
2606B System SourceMeter® Instrument Reference Manual
Section 7: TSP command reference
2606B-901-01 Rev. B / May 2018
7-55
digio.trigger[N].overrun
This attribute returns the event detector overrun status.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute (R)
Yes
Instrument reset
Digital I/O trigger
N
clear
Digital I/O trigger
N
reset
Recall setup
Not saved
Not applicable
Usage
overrun
= digio.trigger[
N
].overrun
overrun
Trigger overrun state (
true
or
false
)
N
Digital I/O trigger line (1 to 14)
Details
If this is
true
, an event was ignored because the event detector was already in the detected state
when the event occurred.
This is an indication of the state of the event detector built into the line itself. It does not indicate if an
overrun occurred in any other part of the trigger model or in any other detector that is monitoring the
event.
Example
overrun = digio.trigger[1].overrun
print(overrun)
If there is no trigger overrun, the following
text is output:
false
Also see
(on page 7-52)
(on page 7-56)
digio.trigger[N].pulsewidth
This attribute describes the length of time that the trigger line is asserted for output triggers.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute (RW)
Yes
Instrument reset
Digital I/O trigger
N
reset
Recall setup
Not saved
10e-6 (10 µs)
Usage
width
= digio.trigger[
N
].pulsewidth
digio.trigger[
N
].pulsewidth =
width
width
The pulse width (seconds)
N
Digital I/O trigger line (1 to 14)