2600AS-900-01 Rev. B / September 2008
Return to
3-17
Series 2600A System SourceMeter
Section 3: Test Script Processor Interaction
Figure 3-8
GUI after running the non-function script (GPIB)
Example script 2: script that uses functions
The script used in this example is in the file named “MeasCurrFunctionScript.tsp.” This script is
downloaded to the Series 2600A by selecting the
Load “Function” Script
option, and then
clicking the
Load Script (GPIB)
command button. The script is downloaded exactly as shown in
the previous example, except the “open file” and “loadscript” commands reflect the different script
name as shown below.
Open App.Path & "\" & "MeasCurrFunctionScript.tsp" For Input As #1
Call send(Addr%, "loadscript MeasCurrFunctionScript", intStatus)
The other difference is in the script itself, which is listed below. This script does not perform
multiple current measurements, but rather it creates a function named “MeasCurr,” which performs
one or more current measurements when it is called. The number of measurements depends on
the value of “ntimes,” which is passed to the function when it is called.
shows the GUI
(Graphical User Interface) after loading and running the script. Note that the
Data
text box does
not contain any readings as in the previous example; it simply indicates that the
“MeasCurr(ntimes)” function has been created.
function MeasCurr(ntimes)
local l_i -- Local counter variable
reset()
display.screen = display.SMUA
display.smua.measure.func = display.MEASURE_DCAMPS
smua.source.func = smua.OUTPUT_DCVOLTS
smua.source.autorangev = smua.AUTORANGE_ON