Series 2600A System SourceMeter® Instruments Reference Manual
Section 19: Remote Commands
Table 19-10
KISweep TSP test script: PulseIMeasureV
TSP project name:
KIPulse
TSP test script name:
KIPulse
Function:
PulseIMeasureV
PulseIMeasureV(smu, bias, level, ton, toff, points)
Description
Performs a specified number of pulse I, measure V cycles:
• Sets the
smu
to output
bias
amps and dwell for
ton
seconds.
• Sets the
smu
to output
level
amps and dwell for
ton
seconds.
• Performs voltage measurement with source at
level
amps.
• Sets the
smu
to output
bias
amps for
toff
seconds.
• Repeats the above sequence for
points
pulse-measure cycles.
Parameters
smu, bias, level, ton, toff, points
smu:
SourceMeter instrument channel (A or B).
Defaults to SMU A if all parameters are omitted
when function is called.
bias:
Bias level in amps.
level:
Pulse level in amps.
ton:
Pulse on time in seconds.
toff:
Pulse off time in seconds.
points:
Number of pulse-measure cycles.
Data
Pulsed voltage measurements, current levels, and timestamps are stored in
smuX.nvbuffer1
.
Example
PulseIMeasureV(smua, 0.001, 1.0, 20E-3, 40E-3, 10)
SMU A will output 1mA and dwell for 20ms, output 1A and dwell for 20ms, and
then perform a voltage measurement. After the measurement, the output will
return to 1mA and dwell for 40ms. This pulse-measure process will repeat 9 more
times.