19-186
Return to
2600AS-901-01 Rev. B / September 2008
Section 19: Remote Commands
Series 2600A System SourceMeter® Instruments Reference Manual
status.operation.trigger_overrun.condition
status.operation.trigger_overrun.enable
status.operation.trigger_overrun.event
status.operation.trigger_overrun.ntr
status.operation.trigger_overrun.ptr
Attribute
Operation status trigger overrun register set.
Default
0
TSP-Link
accessibility
This attribute can be accessed from a remote TSP-Link node.
Usage
Reads condition, enable, event, NTR, and PTR registers:
operreg = status.operation.trigger_overrun.condition
operreg = status.operation.trigger_overrun.enable
operreg = status.operation.trigger_overrun.event
operreg = status.operation.trigger_overrun.ntr
operreg = status.operation.trigger_overrun.ptr
Writes to enable, NTR, and PTR registers:
status.operation.trigger_overrun.enable = operreg
status.operation.trigger_overrun.ntr = operreg
status.operation.trigger_overrun.ptr = operreg
Set
operreg
to one of the following values:
0
Clears all bits.
status.operation.trigger_overrun.SMUA
Sets SMUA bit (B1).
status.operation.trigger_overrun.SMUB
Sets SMUB bit (B2).
status.operation.trigger_overrun.TRIGGER_BLENDER
Sets TRGBLND bit (B10).
status.operation.trigger_overrun.TRGBLND
Sets TRGBLND bit (B10).
status.operation.trigger_overrun.TRIGGER_TIMER
Sets TRGTMR bit (B11).
status.operation.trigger_overrun.TRGTMR
Sets TRGTMR bit (B11).
status.operation.trigger_overrun.DIGITAL_IO
Sets DIGIO bit (B12).
status.operation.trigger_overrun.DIGIO
Sets DIGIO bit (B12).
status.operation.trigger_overrun.TSPLINK
Sets TSPLINK bit (B13).
status.operation.trigger_overrun.LAN
Sets LAN bit (B14).
operreg
can also be set to the decimal weight of the bit to be set. Examples:
To set bit B1 (SMUA), set
operreg
to 2 (2
1
).
To set bit B2 (SMUB), set
operreg
to 4 (2
2
).
To set bit B13 (TSPLINK), set
operreg
to 8192 (2
13
).
To set two bits, set
operreg
to the sum of the decimal weights of both bits:
To set bits B1 and B2, set
operreg
to 6 (2 + 4).