19-174
Return to
2600AS-901-01 Rev. B / September 2008
Section 19: Remote Commands
Series 2600A System SourceMeter® Instruments Reference Manual
status.operation.instrument.smuX.condition
status.operation.instrument.smuX.enable
status.operation.instrument.smuX.event
status.operation.instrument.smuX.ntr
status.operation.instrument.smuX.ptr
smuX = smua or smub
Attribute
Operation status SMU X summary register set.
Default
0
TSP-Link
accessibility
This attribute can be accessed from a remote TSP-Link node.
Usage
Reads condition, enable, event, NTR, and PTR registers:
operreg = status.operation.instrument.smuX.condition
operreg = status.operation.instrument.smuX.enable
operreg = status.operation.instrument.smuX.event
operreg = status.operation.instrument.smuX.ntr
operreg = status.operation.instrument.smuX.ptr
Writes to enable, NTR, and PTR registers:
status.operation.instrument.smuX.enable = operreg
status.operation.instrument.smuX.ntr = operreg
status.operation.instrument.smuX.ptr = operreg
Set
operreg
to one of the following values:
0
status.operation.instrument.smuX.CALIBRATING
status.operation.instrument.smuX.CAL
status.operation.instrument.smuX.SWEEPING
status.operation.instrument.smuX.SWE
status.operation.instrument.smuX.MEASURING
status.operation.instrument.smuX.MEAS
status.operation.instrument.smuX.TRIGGER_OVERRUN
status.operation.instrument.smuX.TRGOVR
Clears all bits.
Sets CAL bit (B0).
Sets CAL bit (B0).
Sets SWE bit (B3).
Sets SWE bit (B3).
Sets MEAS bit (B4).
Sets MEAS bit (B4).
Sets TRGOVR bit (B10).
Sets TRGOVR bit (B10).
operreg
can also be set to the decimal weight of the bit to be set. Examples:
To set bit B0 (CAL), set
operreg
to 1 (2
0
).
To set bit B4 (MEAS), set
operreg
to 16 (2
4
).
To set more than one bit of the register, set
operreg
to the sum of their decimal weights. For
example, to set bits B0 and B4, set
operreg
to 17 (1 + 16).
Remarks
• These attributes are used to read or write to the operation status SMU X summary registers.
• Reading a status register returns a value. The binary equivalent of the returned value indicates
which register bits are set. The least significant bit of the binary number is bit 0, and the most
significant bit is bit 15.
• For example, assume value 17 is returned for the enable register. The binary equivalent is
0000000000010001. This value indicates that bit B0 (CAL) and bit B4 (MEAS) are set.
• The used bits of the operation status SMU X summary registers are described as follows:
•
Bit B0, CAL
: Set bit indicates that SMU X is calibrating.
•
Bit B3, SWE
: Set bit indicates that SMU X is sweeping.
•
Bit B4, MEAS
: Bit will be set when taking an overlapped measurement, but it will not set when
taking a normal synchronous measurement.
•
Bit B10, TRGOVR
: Set bit indicates a bit has been set in the operation status SMU X trigger
overrun event register.
Details
.
Example
Sets the MEAS bit of the operation status SMU A summary enable register:
status.operation.instrument.smua.enable =
status.operation.instrument.smua.MEAS