19-96
Return to
2600AS-901-01 Rev. B / September 2008
Section 19: Remote Commands
Series 2600A System SourceMeter® Instruments Reference Manual
reset
printnumber
Function
Prints numbers using the format selected for printing reading buffers.
TSP-Link
accessibility
This function cannot be accessed from a remote TSP-Link node.
Usage
There are multiple ways to use this function, depending on how many numbers are to be printed:
printnumber(v1)
printnumber(v1 ,v2)
printnumber(v1 ,v2, ..., vn)
v1, v2, ..., vn
Numbers to print.
Remarks
• This function will print the given numbers using the data format specified by
format.data
and
other associated attributes.
• At least one number must be given.
Also see
,
Example
Prints three measurements that were previously performed:
format.data = format.ASCII
printnumber(i, v, t)
Example of returned data (i, v, t):
1.02345E-04, 8.76542E-02, 5.29372E-01
reset
Function
Resets the logical instruments to the default settings.
TSP-Link
accessibility
This function cannot be accessed from a remote TSP-Link node.
Usage
reset()
reset(system)
system
Flag indicating what part of the system to reset. Can
be set to
true
or
false
.
Remarks
• If
system
is set to
true
and the node is the master, the entire system will be reset.
• If
system
is set to
false
, only the local group will be reset.
• The default value of
system
when no value is given is
true
.
• This function resets all logical instruments in the system or local group. This function is
equivalent to iterating over all the logical instruments and calling the reset method of each.
• Resetting the entire system is only permitted when the node is the master. If the node is not the
master, it generates an error.
Also see