19-94
Return to
2600AS-901-01 Rev. B / September 2008
Section 19: Remote Commands
Series 2600A System SourceMeter® Instruments Reference Manual
meminfo
opc
This function sets the OPC bit in the status register when all overlapped commands are
completed.
meminfo
Function
Returns the current amount of available memory and the total amount of memory in the instrument.
TSP-Link
accessibility
This function cannot be accessed from a remote TSP-Link node.
Usage
freemem, totalmem = meminfo()
freemem
Amount of free dynamically allocated memory available.
totalmem
The total amount of dynamically allocated memory
in the instrument.
Remarks
• This function returns two values:
•
The amount of free dynamically allocated memory available in kilobytes.
•
The total amount of dynamically allocated memory on the instrument in kilobytes.
• The difference between the two is the amount currently used.
opc
Function
Sets the Operation Complete status bit when all overlapped commands are completed.
TSP-Link
accessibility
This function cannot be accessed from a remote TSP-Link node.
Usage
opc()
Remarks
• This function will cause the Operation Complete bit in the Standard Event Status Register to be
set when all previously started local overlapped commands are complete. Note that each node
will independently set their Operation Complete bits in their own status models.
• Any nodes not actively performing overlapped commands will set their bits immediately. All
remaining nodes will set their own bits as they complete their own overlapped commands.
Details
See