Series 2600A System SourceMeter® Instruments Reference Manual
Section 19: Remote Commands
Example
Prompts the operator to enter a voltage value – Valid input range is 0 to +2.00, with a default of
0.50:
value = display.prompt("0.00", "V", "Input 0 to +2V", 0.5, 0, 2)
The above command will display the following input prompt:
0.50V
Input 0 to +2V
display.screen
Attribute
The selected display screen.
Default
display.SMUA
(2601A/2611A/2635A)
display.SMUA_SMUB
(2602A/2612A/2636A)
TSP-Link
accessibility
This attribute can be accessed from a remote TSP-Link node.
Usage
displayid = display.screen
-- Reads display screen.
display.screen = displayid
-- Writes display screen.
Set
displayid
to one of the following values:
0
or
display
.
SMUA
Displays source-measure and compliance for SMU A.
1
or
display
.
SMUB
Displays source-measure and compliance for SMU B.
2
or
display.SMUA_SMUB
Displays source-measure for SMU A and SMU B.
3
or
display.USER
Displays the user screen.
Remarks
• Setting this attribute selects the display screen for the front panel. This attribute can be read to
determine which of the four available display screens was last selected by the user. The user
can select the screen by value or one of the enumerations.
Details
See
in
.
Example
Selects the source-measure and compliance limit display for SMUA:
display.screen = display.SMUA