2600AS-901-01 Rev. B / September 2008
Return to
12-13
Series 2600A System SourceMeter® Instruments Reference Manual
Section 12: TSP Fundamentals and Script Management
Loading scripts from the USB flash drive
You can use the
LOAD
feature to load a script from the USB flash drive to the run time
environment. The Series 2600A validates the script before the load is completed. You can view the
errors on the front panel of the Series 2600A.
Once a script is loaded from the USB flash drive, you can associate the script with the front panel's
RUN
button or save the script to internal nonvolatile memory.
ACTIVE-FOR-RUN:
If this option is selected, the script can be executed by pressing the front
panel's
RUN
button.
SAVE-INTERNAL:
If this option is selected, the script is saved to internal nonvolatile memory.
To load a script from the USB flash drive:
1.
Press
MENU
>
SCRIPT
and then choose
LOAD
.
2.
Choose
USB1
.
3.
Turn the
navigation wheel
left or right to view the files on the USB flash drive.
4.
Select the desired file by highlighting it and pressing
ENTER
or the
navigation wheel
.
5.
Choose one of the following:
•
(Save only)
SAVE-INTERNAL
•
Select
YES
.
•
(Run only)
ACTIVE-FOR-RUN
.
•
Select
YES
.
6.
If you chose
ACTIVE-FOR-RUN
, execute the script by pressing
EXIT
and then
RUN
.
Working with subdirectories from the front panel
To access subdirectories while in local mode:
1.
From the front panel, press
MENU
>
SCRIPT
>
LOAD
>
USB1
.
2.
Turn the
navigation wheel
left or right to select the desired directory. The files in the
subdirectory are displayed.
Figure 12-5
Subdirectories