9-4
Return to
2600AS-901-01 Rev. B / September 2008
Section 9: Sweep Operation
Series 2600A System SourceMeter® Instruments Reference Manual
Figure 9-2
Linear staircase sweep
A linear staircase sweep is configured using a start level, a stop level, and the total number of
points, including the start and stop points. The step size is determined by the start and stop levels,
and the number of sweep points:
step = (stop - start) / (points - 1)
NOTE The number of sweep steps actually performed is determined by the
trigger count. Refer to
for more information.
The sweep can be either positive-going or negative-going, depending on the relative values of the
start and stop parameters. When the sweep starts, the output will go to the start source level. The
output will then change in equal steps until the stop level is reached. If the trigger count is greater
than the number of points specified, the SMU will start over at the beginning value.
To configure a linear staircase sweep, use the following function:
smuX.trigger.source.linearY
This function configures the source values the SMU will output when performing a linear sweep.
After configuring the sweep you must also enable the source action by setting the following
attribute:
smuX.trigger.source.action
Example:
-- Sweep from 0 to 10V in 1V steps.
smua.trigger.source.linearv(0, 10, 11)
-- Enable the source action.
smua.trigger.source.action = smua.ENABLE
For more information, see
.
Delay
Delay
Delay
Delay
X = Measurement
X
X
X
X
Measure
Measure
Measure
Start
Step
Step
Step
Stop
Measure