Model 2520INT Quick Start Guide
Getting Started
2-7
Programming example
Table 2-2 summarizes the commands for the 2520INT connected to Detector 1. These
commands set up the Model 2520 as follows:
•
Measurement range: 10mA
•
Bias Voltage: -5V
•
Measurement polarity: Positive
CAUTION
Applying voltage greater than -5V or .5V can damage the
Model 2520INT.
Detector dark current measurements
Description
The dark current is a current offset produced by the detector in the absence of light. Its
value is based on many factors including the reverse bias voltage on the detector, tempera-
ture, semiconductor architecture, and even the quality of the semiconductor material itself.
Proper measurement of and compensation for the dark current will ensure accurate optical
power measurements.
Table 2-1
Remote voltage bias and detector polarity measurement commands to bias 2520INT
Command
Description
:SOURce2:VOLTage <voltage>
Set detector #1 source voltage (0 to ±20).
:SOURce3:VOLTage <voltage>
Set detector #2 source voltage (0 to ±20).
:SENSe2:CURRent:RANGe <range>
Set detector #1 measure range (0.01, 0.02, 0.05, 0.1).
:SENSe2:CURRent:POLarity <polarity>
Set detector #1 measure polarity.
:SENSe3:CURRent:RANGe <range>
Set detector #2 measure range (0.01, 0.02, 0.05, 0.1).
:SENSe3:CURRent:POLarity <polarity>
Set detector #2 measure polarity.
Table 2-2
Command summary for 2520INT connected to Detector 1
Command
Description
:SENS2:CURR:RANG 0.01
Detector 1 measurement range at 10mA
:SENS2:CURR:POL POS
Detector 1 measurement polarity positive
:SOUR2:VOLT –5
Detector 1 bias = -5V