F-10
Measurement Considerations
Model 2520 User’s Manual
Magnetic coupling
Magnetic coupling between sense and current leads can also affect measurements.
Figure F-9A shows connections that will result in considerable coupling due to the rela-
tively long exposed signal lead area. Carrying the sense shields as close to the DUT as
possible (Figure F-9B) will reduce magnetic coupling and improve response. Twisting the
exposed sense leads and even the cables themselves may be necessary to reduce coupling
to an acceptable level.
Figure F-9
Magnetic coupling
DUT
Connecting
Cables
Coupling
Caused by
Exposed Leads
A. Coupling caused by exposed leads.
DUT
Connecting
Cables
Coupling
Reduced by
Shielding
B. Coupling reduced by shielding.
Voltage Sense HI
Current Output HI
Current Output LO
Voltage Sense LO
Voltage Sense HI
Current Output HI
Current Output LO
Voltage Sense LO
Содержание 2520
Страница 250: ...A Specifications ...
Страница 254: ...B StatusandErrorMessages ...
Страница 262: ...C DataFlow ...
Страница 266: ...D IEEE 488BusOverview ...
Страница 281: ...E IEEE 488andSCPI ConformanceInformation ...
Страница 285: ...F MeasurementConsiderations ...
Страница 308: ...G GPIB488 1Protocol ...
Страница 312: ...H ExamplePrograms ...
Страница 317: ...I ContinuousPulseMode ...