Model 2520 User’s Manual
Measurement Considerations
F-9
Figure F-8 shows response of a 2A pulse with voltage sense leads 1/4 inch away from the
DUT. Note the long settling tail corresponding to the DUT resistance (1.66
Ω
) and the
inductance formed by the leads.
Figure F-8
Response of 2A pulse with sense leads 1/4 inch away from DUT
2.5
0.8
2.0
1.5
1.0
0.5
0.0
0
1
1.6
1.8
2.2
Current
(A)
Time (
µ
s)
0.6
1.2
1.4
2
2.4
Содержание 2520
Страница 250: ...A Specifications ...
Страница 254: ...B StatusandErrorMessages ...
Страница 262: ...C DataFlow ...
Страница 266: ...D IEEE 488BusOverview ...
Страница 281: ...E IEEE 488andSCPI ConformanceInformation ...
Страница 285: ...F MeasurementConsiderations ...
Страница 308: ...G GPIB488 1Protocol ...
Страница 312: ...H ExamplePrograms ...
Страница 317: ...I ContinuousPulseMode ...