Model 2520 User’s Manual
SCPI Command Reference
14-35
Query
:LOW?
Query programmed low amplitude
:LOW? DEFault
Query *RST default low amplitude
:LOW? MINimum
Query lowest allowable low amplitude
:LOW? MAXimum
Query highest allowable low amplitude
Description
This command is used to set the pulse low amplitude level for all current
source pulses. For the 500mA source range, the low value can be set
from 0 to 0.015A. For the 5A source range, the low value can be set
from 0 to 0.150A.
NOTE
This command sets the pulse low amplitude for the pulse source function in the
list and sweep modes. See
“Select sourcing mode,” page 14-32
.
Set voltage limit
PROTection[:LEVel] <NRf>
:SOURce[1]:VOLTage:PROTection[:LEVel] <NRf>
Set voltage compliance limit
Parameters
<NRf> =
3 to 10.5
Set voltage compliance limit
Query
:PROTection?
Query programmed voltage limit
Description
This command is used to the set the voltage protection limit, which
clamps the current source output voltage at the programmed limit. See
, “
Current source operating boundaries
,
”
for more information.
Select source polarity
POLarity <name>
[:SOURce[1]]:CURRent:POLarity <name>
Set laser diode current source polarity
Parameters
<name> =
POSitive
Positive polarity
NEGative
Negative polarity
Query
:POLarity?
Query measurement polarity
Description
This command is used to select the polarity for the laser diode current
source. POSitive selects normal polarity, while NEGative selects
reversed polarity.
See
, “
,”
for more information on
polarity.
Содержание 2520
Страница 250: ...A Specifications ...
Страница 254: ...B StatusandErrorMessages ...
Страница 262: ...C DataFlow ...
Страница 266: ...D IEEE 488BusOverview ...
Страница 281: ...E IEEE 488andSCPI ConformanceInformation ...
Страница 285: ...F MeasurementConsiderations ...
Страница 308: ...G GPIB488 1Protocol ...
Страница 312: ...H ExamplePrograms ...
Страница 317: ...I ContinuousPulseMode ...