Model 2520 User’s Manual
Digital I/O Port, Interlocks, and Pulse Sync Output
9-7
Interlock status indicator test sequence
Perform the following steps to verify the operation of the indicator lights:
1.
Remove any connections from the testhead CURRENT OUTPUT, VOLTAGE
SENSE, or DETECTOR terminals.
2.
Connect the testhead to the mainframe, and follow the steps for power connection
and power-up sequence in
3.
Bypass the interlock switch on the remote testhead with a bypass plug or cable,
connected as shown in
4.
Insert the key provided into the key interlock, and verify it is in the DISABLED
position.
At this point the INTERLOCK STATUS indicator should be BLUE, and the
LASER POWER ON indicator should be OFF on the testhead. The OUTPUT
ON/OFF indicator on the mainframe should be OFF.
5.
Turn the key in the key interlock to ENABLED.
At this point the INTERLOCK STATUS indicator should be GREEN, and the
LASER POWER ON indicator should be OFF on the testhead. The OUTPUT ON/
OFF indicator on the mainframe should be OFF.
6.
Press the OUTPUT ON/OFF switch on the mainframe.
At this point the INTERLOCK STATUS indicator should be GREEN, and the
LASER POWER ON indicator should be BLUE on the testhead. The OUTPUT
ON/OFF indicator on the mainframe should be BLUE.
7.
Turn the key in the key interlock to DISABLED.
At this point the INTERLOCK STATUS indicator should be BLUE, and the
LASER POWER ON indicator should be OFF on the testhead. The OUTPUT
ON/OFF indicator on the mainframe should be OFF.
WARNING
If at any time the indicators provided on the testhead, for INTER-
LOCK STATUS or LASER POWER ON, should fail to light or to
properly indicate status, immediately contact a Keithley service repre-
sentative for repair. Failure to do so may expose the user to hazards
without proper warnings.
Reading interlock state
To read the state of the interlocks, send this query via remote:
:OUTP1:INT:TRIP?
The response to this query is either 0 or 1 as follows:
•
A returned value of 1 indicates that at least one of the interlocks is tripped (circuit
open), and the outputs cannot be turned on.
•
A returned value of 0 indicates neither interlock is tripped (circuit closed), and the
outputs can be turned on.
Содержание 2520
Страница 250: ...A Specifications ...
Страница 254: ...B StatusandErrorMessages ...
Страница 262: ...C DataFlow ...
Страница 266: ...D IEEE 488BusOverview ...
Страница 281: ...E IEEE 488andSCPI ConformanceInformation ...
Страница 285: ...F MeasurementConsiderations ...
Страница 308: ...G GPIB488 1Protocol ...
Страница 312: ...H ExamplePrograms ...
Страница 317: ...I ContinuousPulseMode ...