Model 2520 Pulsed Laser Diode Test System Service Manual
Troubleshooting
4-17
No comm link error
A “No Comm Link” error indicates that the front panel display processor has stopped
communicating with the main processor, which is located on the motherboard. This error
indicates that the main processor ROMs (U3 and U4) may require re-seating in their sock-
ets. The ROMs may be reseated as follows:
1. Turn off the power, and disconnect the line cord and all other test leads and cables
from the instrument.
2. Remove the case cover as outlined in
“Case cover removal” on page 5-4
.
3. Locate the firmware ROMs, U3 and U4, located on the motherboard. These ROMs
are the only IC installed in a socket. (Refer to the component layout drawing at the
end of
for exact location.)
4. Carefully push down on the ROM ICs to make sure it is properly seated in its
socket.
CAUTION
Be careful not to push down excessively. The motherboard could
crack.
5. Connect the line cord, and turn on the power. If the problem persists, additional
troubleshooting will be required.
Содержание 2520
Страница 10: ......
Страница 13: ...1 PerformanceVerification...
Страница 30: ......
Страница 31: ...2 Calibration...
Страница 68: ...2 38 Calibration Model 2520 Pulsed Laser Diode Test System Service Manual...
Страница 69: ...3 RoutineMaintenance...
Страница 72: ...3 4 Routine Maintenance Model 2520 Pulsed Laser Diode Test System Service Manual...
Страница 73: ...4 Troubleshooting...
Страница 90: ...4 18 Troubleshooting Model 2520 Pulsed Laser Diode Test System Service Manual...
Страница 91: ...5 Disassembly...
Страница 98: ...5 8 Disassembly Model 2520 Pulsed Laser Diode Test System Service Manual...
Страница 99: ...6 ReplaceableParts...
Страница 116: ...6 18 Replaceable Parts Model 2520 Pulsed Laser Diode Test System Service Manual...
Страница 117: ...A Specifications...
Страница 121: ...B CalibrationReference...
Страница 132: ...B 12 Calibration Reference Model 2520 Pulsed Laser Diode Test System Service Manual...
Страница 136: ......