1-12
Performance Verification
Model 2520 Pulsed Laser Diode Test System Service Manual
Figure 1-5
Resistor mounting and connections
Resistor Edge
View
0.5in
2
Maximum Loop
Area
Resistors
Board
9. Calculate the current using the voltage value and characterized resistance value:
I = V/R. Recalculate reading limits from these values, then verify that the current is
within those limits.
10. Replace the 10
Ω
to 15
Ω
resistor with the characterized 1.2
Ω
to 1.6
Ω
resistor.
11. Repeat steps 5 through 9 for the 5A range using the 1A value shown in
12. For pulse mode verification, change connections as shown in
repeat the above procedure. for the pulse mode.
Содержание 2520
Страница 10: ......
Страница 13: ...1 PerformanceVerification...
Страница 30: ......
Страница 31: ...2 Calibration...
Страница 68: ...2 38 Calibration Model 2520 Pulsed Laser Diode Test System Service Manual...
Страница 69: ...3 RoutineMaintenance...
Страница 72: ...3 4 Routine Maintenance Model 2520 Pulsed Laser Diode Test System Service Manual...
Страница 73: ...4 Troubleshooting...
Страница 90: ...4 18 Troubleshooting Model 2520 Pulsed Laser Diode Test System Service Manual...
Страница 91: ...5 Disassembly...
Страница 98: ...5 8 Disassembly Model 2520 Pulsed Laser Diode Test System Service Manual...
Страница 99: ...6 ReplaceableParts...
Страница 116: ...6 18 Replaceable Parts Model 2520 Pulsed Laser Diode Test System Service Manual...
Страница 117: ...A Specifications...
Страница 121: ...B CalibrationReference...
Страница 132: ...B 12 Calibration Reference Model 2520 Pulsed Laser Diode Test System Service Manual...
Страница 136: ......