Model 2461 Interactive SourceMeter® Instrument Reference Manual
Section 8: TSP command reference
2461-901-01 A/November 2015
8-305
To use the timer as a simple delay or pulse generator with digital I/O lines, make sure the timer start
time in seconds and fractional seconds is configured for a time in the past. To use the timer as an
alarm, configure the timer start time in seconds and fractional seconds for the desired alarm time.
Example
trigger.timer[3].enable = trigger.ON
Enable the trigger timer for timer 3.
Also see
None
trigger.timer[N].reset()
This function resets trigger timer settings to their default values.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Function
Yes
Usage
trigger.timer[
N
].reset()
N
Trigger timer number (1 to 4)
Details
The
trigger.timer[
N
].reset()
function resets the following attributes to their default values:
•
trigger.timer[
N
].count
•
trigger.timer[
N
].delay
•
trigger.timer[
N
].delaylist
•
trigger.timer[
N
].enable
•
trigger.timer[
N
].start.generate
•
trigger.timer[
N
].start.fractionalseconds
•
trigger.timer[
N
].start.seconds
•
trigger.timer[
N
].stimulus
It also clears
trigger.timer[
N
].overrun
.
Example
trigger.timer[1].reset()
Resets the attributes associated with timer 1 to
their default values.
Also see
(on page 8-301)
(on page 8-303)
(on page 8-303)
(on page 8-304)
trigger.timer[N].start.fractionalseconds
(on page 8-306)
trigger.timer[N].start.generate
trigger.timer[N].start.overrun
(on page 8-307)
trigger.timer[N].start.seconds
(on page 8-308)
trigger.timer[N].start.stimulus
(on page 8-308)