Model 2461 Interactive SourceMeter® Instrument Reference Manual
Section 8: TSP command reference
2461-901-01 A/November 2015
8-197
smu.source.pulsesweeplinearstep()
This function sets up a linear source pulse sweep configuration list model with a fixed number of steps.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Function
Yes
Usage
smu.source.pulsesweeplinearstep(
configListName, biasLevel, start, stop, step,
pulseWidth)
smu.source.pulsesweeplinearstep(
configListName, biasLevel, start, stop, step,
pulseWidth, measEnable)
smu.source.pulsesweeplinearstep(
configListName, biasLevel, start, stop, step,
pulseWidth, measEnable, bufferName)
smu.source.pulsesweeplinearstep(
configListName, biasLevel, start, stop, step,
pulseWidth, measEnable, bufferName, sDelay)
smu.source.pulsesweeplinearstep(
configListName, biasLevel, start, stop, step,
pulseWidth, measEnable, bufferName, sDelay, offTime)
smu.source.pulsesweeplinearstep(
configListName, biasLevel, start, stop, step,
pulseWidth, measEnable, bufferName, sDelay, offTime, count)
smu.source.pulsesweeplinearstep(
configListName, biasLevel, start, stop, step,
pulseWidth, measEnable, bufferName, sDelay, offTime, count, xBiasLimit)
smu.source.pulsesweeplinearstep(
configListName, biasLevel, start, stop, step,
pulseWidth, measEnable, bufferName, sDelay, offTime, count, xBiasLimit,
xPulseLimit)
smu.source.pulsesweeplinearstep(
configListName, biasLevel, start, stop, step,
pulseWidth, measEnable, bufferName, sDelay, offTime, count, xBiasLimit,
xPulseLimit, failAbort)
smu.source.pulsesweeplinearstep(
configListName, biasLevel, start, stop, step,
pulseWidth, measEnable, bufferName, sDelay, offTime, count, xBiasLimit,
xPulseLimit, failAbort, dual
)
configListName
The name of the source configuration list that the sweep uses; this must be defined
before sending this command
biasLevel
Output level the instrument sources before the first pulse and returns to between
pulses (delay and off time):
•
Current: -7.35 A to 7.35 A
•
Voltage: -105 V to 105 V
start
The voltage or current source level at which the pulse sweep starts:
•
Current: -10.5 A to 10.5 A
•
Voltage: -105 V to 105 V
stop
The voltage or current source level at which the pulse sweep stops:
•
Current: -10.5 A to 10.5 A
•
Voltage: -105 V to 105 V
step
The step size at which the source level will change; step size must be greater than
0; to calculate the number of source-measure points in a sweep, use the following
formula:
Points = [(Stop - Start) / Step] + 1
pulseWidth
The time at the amplitude level for each pulse:
•
Extended operating area: 10 A at 100 V, 5 % duty cycle, maximum pulse
width: 1 ms, minimum pulse width 150 µs (load dependent)
•
Normal operating area (DC): 99.99 % duty cycle, maximum pulse width
10,000 s, minimum pulse width 150 µs