Section 8: TSP command reference
Model 2461 Interactive SourceMeter® Instrument Reference Manual
8-138
2461-901-01 A/November 2015
smu.measure.configlist.query()
This function returns a list of TSP commands and parameter settings that are stored in the specified configuration
index.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Function
Yes
Usage
smu.measure.configlist.query(
listName
,
index
)
smu.measure.configlist.query(
listName
,
index, fieldSeparator
)
listName
A string that represents the name of a measure configuration list
index
A number that defines a specific configuration index in the configuration list
fieldSeparator
String that represents the separator for the data; use one of the following:
•
Comma (default):
,
•
Semicolon:
;
•
New line:
\n
Details
This command recalls data for one configuration index from the specified measure configuration list
and from the source configuration list (if specified).
For additional information about the information this command recalls when using a configuration list
query command, see
Instrument settings stored in a measure configuration list