Model 2461 Interactive SourceMeter® Instrument Reference Manual
Section 8: TSP command reference
2461-901-01 A/November 2015
8-103
smu.contact.checkall()
This function checks the high, low, and guard connections for excessive contact resistance.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Function
Yes
Usage
high, low, guard
= smu.contact.checkall()
high
The result of the contact check operation for the high connection:
•
true
: The connection did not exceed the threshold resistance value
•
false
: The connection exceeded the threshold resistance value
low
The result of the contact check operation for the low connection:
•
true
: The connection did not exceed the threshold resistance value
•
false
: The connection exceeded the threshold resistance value
guard
The result of the contact check operation for the guard connection:
•
true
: The connection did not exceed the threshold resistance value
•
false
: The connection exceeded the threshold resistance value
Details
Use this command to check the high, low, and guard connections to the device under test (DUT) for
excessive contact resistance.
Use the
print()
command to query the result of the contact check operation. For each of the three
connections, this command returns
false
if the threshold resistance exceeds the level specified by
the
smu.contact.threshold
command, or
true
if it does not exceed that value.
If you get a failed indication when sending the
smu.contact.check()
function, use this function to
determine which connection failed.
Sending this command when contact check is not enabled results in an error.
Example
smu.contact.enable = smu.ON
print(smu.contact.checkall())
Enable contact check.
Get the status of high, low, and guard
connections for contact check.
Output:
false false true
Indicates that the high and low connections
exceeded the specified contact check resistance
threshold value, but the guard connection did not
exceed that value.
Also see
(on page 8-102)
(on page 8-104)
(on page 8-105)