Model 2461 Interactive SourceMeter® Instrument Reference Manual
Section 6: SCPI command reference
2461-901-01 A/November 2015
6-121
:SOURce[1]:PULSe:TRain:<function>
This command defines a sequence of source pulses and creates a trigger model to generate the pulse train.
Type
Affected by
Where saved
Default value
Command only
Recall settings
Instrument reset
Power cycle
Save settings
Source configuration list
Measure configuration list
Not applicable
Usage
:SOURce[1]:PULSe:TRain:<function> <biasLevel>, <pulseLevel>, <pulseWidth>, <count>
:SOURce[1]:PULSe:TRain:<function> <biasLevel>, <pulseLevel>, <pulseWidth>, <count>,
<measEnable>
:SOURce[1]:PULSe:TRain:<function> <biasLevel>, <pulseLevel>, <pulseWidth>, <count>,
<measEnable>, "<bufferName>"
:SOURce[1]:PULSe:TRain:<function> <biasLevel>, <pulseLevel>, <pulseWidth>, <count>,
<measEnable>, "<bufferName>", <delay>
:SOURce[1]:PULSe:TRain:<function> <biasLevel>, <pulseLevel>, <pulseWidth>, <count>,
<measEnable>, "<bufferName>", <delay>, <offTime>
:SOURce[1]:PULSe:TRain:<function> <biasLevel>, <pulseLevel>, <pulseWidth>, <count>,
<measEnable>, "<bufferName>", <delay>, <offTime>, <xBiasLimit>
:SOURce[1]:PULSe:TRain:<function> <biasLevel>, <pulseLevel>, <pulseWidth>, <count>,
<measEnable>, "<bufferName>", <delay>, <offTime>, <xBiasLimit>, <xPulseLimit>
:SOURce[1]:PULSe:TRain:<function> <biasLevel>, <pulseLevel>, <pulseWidth>, <count>,
<measEnable>, "<bufferName>", <delay>, <offTime>, <xBiasLimit>, <xPulseLimit>,
<failAbort>
<function>
The source function:
•
Voltage pulse sweep:
VOLTage
•
Current pulse sweep:
CURRent
<biasLevel>
Output level the instrument sources before the first pulse and returns to between
pulses (delay and off time):
•
Current: -7.35 A to 7.35 A
•
Voltage: -105 V to 105 V
<pulseLevel>
The amplitude current or voltage from zero (not from the bias level):
•
Current: -10.5 A to 10.5 A
•
Voltage: -105 V to 105 V
<pulseWidth>
The time at the amplitude level for each pulse:
•
Extended operating area: 10 A at 100 V, 5 % duty cycle, maximum pulse
width: 1 ms, minimum pulse width 150 µs (load dependent)
•
Normal operating area (DC): 99.99 % duty cycle, maximum pulse width
10,000 s, minimum pulse width 150 µs
<count>
The number of pulses in the pulse train; default is
1
:
•
Infinite loop:
0
•
Finite loop:
1
to
268435455
<measEnable>
Enable or disable measurements at the top of each pulse:
•
Enable:
ON
(default)
•
Disable:
OFF
<bufferName>
A string that indicates the reading buffer; the default buffers (
defbuffer1
or
defbuffer2
) or the name of a user-defined buffer; if no buffer is specified,
defbuffer1
is used
<delay>
The amount of time that the instrument stays at bias level before each pulse (in
seconds): 0 to 10,000 s