Section 6: SCPI command reference
Model 2461 Interactive SourceMeter® Instrument Reference Manual
6-116
2461-901-01 A/November 2015
:SOURce[1]:PULSe:SWEep:<function>:LIST
This command sets up a pulse sweep based on a configuration list, which allows you to customize the sweep.
Type
Affected by
Where saved
Default value
Command only
Recall settings
Instrument reset
Power cycle
Save settings
Not applicable
Usage
SOURce[1]:PULSe:SWEep:<function>:LIST <pulseWidth>
SOURce[1]:PULSe:SWEep:<function>:LIST <pulseWidth>, <MeasEnable>
SOURce[1]:PULSe:SWEep:<function>:LIST <pulseWidth>, <MeasEnable>, "<bufferName>"
SOURce[1]:PULSe:SWEep:<function>:LIST <pulseWidth>, <MeasEnable>, "<bufferName>",
<startIndex>
SOURce[1]:PULSe:SWEep:<function>:LIST <pulseWidth>, <MeasEnable>, "<bufferName>",
<startIndex>, <count>
SOURce[1]:PULSe:SWEep:<function>:LIST <pulseWidth>, <MeasEnable>, "<bufferName>",
<startIndex>, <count>, <delay>
SOURce[1]:PULSe:SWEep:<function>:LIST <pulseWidth>, <MeasEnable>, "<bufferName>",
<startIndex>, <count>, <delay>, <offTime>
SOURce[1]:PULSe:SWEep:<function>:LIST <pulseWidth>, <MeasEnable>, "<bufferName>",
<startIndex>, <count>, <delay>, <offTime>, <failAbort>
SOURce[1]:PULSe:SWEep:<function>:LIST <pulseWidth>, <MeasEnable>, "<bufferName>",
<startIndex>, <count>, <delay>, <offTime>, <failAbort>, <configListName>
<function>
The source function:
•
Voltage pulse sweep:
VOLTage
•
Current pulse sweep:
CURRent
<pulseWidth>
The time at the amplitude level for each pulse:
•
Extended operating area: 10 A at 100 V, 5 % duty cycle, maximum pulse
width: 1 ms, minimum pulse width 150 µs (load dependent)
•
Normal operating area (DC): 99.99 % duty cycle, maximum pulse width
10,000 s, minimum pulse width 150 µs
<measEnable>
Enable or disable measurements at the top of each pulse:
•
Enable:
ON
(default)
•
Disable:
OFF
<bufferName>
A string that indicates the reading buffer; the default buffers (
defbuffer1
or
defbuffer2
) or the name of a user-defined buffer; if no buffer is specified,
defbuffer1
is used
<startIndex>
The index in the configuration list where the pulse sweep starts; default is
1
<count>
The number of pulse sweeps; default is
1
:
•
Infinite loop:
0
•
Finite loop:
1
to
268435455
<delay>
The amount of time that the instrument stays at bias level before each pulse (in
seconds): 0 to 10,000 s
<offTime>
The amount of time that the instrument stays at bias level after each pulse (in
seconds): 0 to 10,000 s
<failAbort>
Determines if the sweep is stopped immediately if a limit is exceeded; options are:
•
Abort the sweep if the source bias or pulse limit is exceeded:
ON
•
Complete the sweep if the source limit is exceeded:
OFF
<configlistName>
The name of the source configuration list that the sweep uses; this must be defined
before sending this command