Model 2461 Interactive SourceMeter® Instrument Reference Manual
Section 6: SCPI command reference
2461-901-01 A/November 2015
6-51
SCRipt subsystem
The SCRipt subsystem controls macro or instrument setup scripts. For additional information on
macro scripts, refer to
Saving front-panel settings into a macro script
(on page 3-29).
SCRipt:RUN
This command runs a script.
Type
Affected by
Where saved
Default value
Command only
Not applicable
Not applicable
Not applicable
Usage
SCRipt:RUN "<scriptName>"
<scriptName>
The name of the script
Details
The script must be available in the instrument to be used by this command.
Example
SCR:RUN "bufferCreate"
Runs a script named
bufferCreate
.
Also see
Saving front-panel settings into a macro script
(on page 3-29)
(on page 2-39)