Model 2461 Interactive SourceMeter® Instrument Reference Manual
Section 6: SCPI command reference
2461-901-01 A/November 2015
6-25
Example
:CALC:VOLT:MATH:FORM MXB
:CALC:VOLT:MATH:MMF 0.80
:CALC:VOLT:MATH:MBF 50
:CALC:VOLT:MATH:STAT ON
Set the math function for voltage measurements to mx+b.
Set the scale factor for voltage measurements to 0.80.
Set the offset factor to 50.
Enable the math function.
Also see
:CALCulate[1]:<function>:MATH:FORMat
(on page 6-19)
Calculations that you can apply to measurements
(on page 3-82)
:CALCulate2:<function>:LIMit<Y>:AUDible
This command determines if the instrument beeper sounds when a limit test passes or fails, or disables the
beeper.
Type
Affected by
Where saved
Default value
Command and query
Recall settings
Instrument reset
Power cycle
Measure configuration list
Save settings
Measure configuration list
NONE
Usage
:CALCulate2:<function>:LIMit<Y>:AUDible <state>
:CALCulate2:<function>:LIMit<Y>:AUDible?
<function>
The measure function:
•
Current:
CURRent[:DC]
•
Resistance:
RESistance
•
Voltage:
VOLTage[:DC]
The digitize function:
•
Digitize current:
DIGitize:CURRent
•
Digitize voltage:
DIGitize:VOLTage
<Y>
Limit number:
1
or
2
<state>
When the beeper sounds:
•
Never:
NONE
•
On test failure:
FAIL
•
On test pass:
PASS
Details
The tone and length of beeper cannot be adjusted.
If you send this command without the
<function>
parameter, it will set the limit for all measure
functions. It will not change the setting for a digitize function.