Section 6: SCPI command reference
Model 2461 Interactive SourceMeter® Instrument Reference Manual
6-12
2461-901-01 A/November 2015
:READ:DIGitize?
This query makes a digitize measurement, places it in a reading buffer, and returns the latest reading.
Type
Affected by
Where saved
Default value
Query only
Not applicable
Not applicable
Not applicable
Usage
:READ:DIGitize?
:READ:DIGitize? "<bufferName>"
:READ:DIGitize? "<bufferName>", <bufferElements>
<bufferName>
The name of the buffer where the reading is stored; if nothing is specified,
defbuffer1
is used
<bufferElements>
See
Details
; if nothing is specified,
READing
is used
Details
You must set the instrument to a digitize function before sending this command.
This query makes the number of readings specified by
[:SENSe[1]]:DIGitize:COUNt
. If multiple
readings are made, all readings are available in the reading buffer. However, only the last reading is
returned as a reading with the command. To get multiple readings, use the
:TRACe:DATA?
command.
When specifying buffer elements, you can:
•
Specify buffer elements in any order.
•
Include any or all of the buffer elements listed below in a single list. You can repeat elements as
long as the number of elements in the list is less than 14.
•
Use a comma to delineate multiple elements for a data point.