Basic Power Supply Operation 2-13
NOTE
Refer to the Programming syntax paragraph of Section 6 for a description of
parameters (e.g., <b>, <NRf>, etc.).
Command notes (outputting voltage and current)
SENSe[1]:CURRent:RANGe <n>
Applies to battery channel (#1)
SENSe2:CURRent:RANGe <n>
Applies to charger channel (#2)
After specifying a current value, the instrument will go to the most sensitive range to
accommodate that reading. For example, if you are expecting a maximum current
reading of 750mA, you can let <n> = 0.75 (or 750e-3) to select the 5A range. Using the
:RANGe command to manually select a current range disables auto range. Another way
to select a range is to use the MINimum, MAXimum, and DEFault parameters as follows:
SENS:CURR:RANG MIN
Select the low current range (5mA) for battery
channel (#1).
SENS2:CURR:RANG MAX
Select the high current range (5A) for charger
channel (#2).
SENS2:CURR:RANG DEF
Select the default current range for charger channel
(#2).
The response for :RANGe? query returns the selected range value which is either 5.0000
or 0.0050.
SENSe[1]:CURRent:RANGe:AUTO <b>
Applies to battery channel (#1)
SENSe2:CURRent:RANGe:AUTO <b>
Applies to charger channel (#2)
This command is coupled to the :RANGe <n> command. When auto range is enabled,
the response for :RANGe? query returns the selected range value which is either 5.0000
or 0.0050. If you then disable auto range, the instrument will remain at the last selected
range.
[SOURce1]:VOLTage <n>
Applies to battery channel (#1)
SOURce2:VOLTage <n>
Applies to charger channel (#2)
This command sets voltage amplitude in volts: 0 to 15 (1mV resolution).
[SOURce1]:CURRent <n>
Applies to battery channel (#1)
SOURce2:CURRent <n>
Applies to charger channel (#2)
• With the 5mA measurement range selected, the maximum current limit is 1A.
• Sending a value that exceeds 1A is rejected, and the following message is displayed
briefly:
CURRENT LIMIT ON
mA RANGE
≤
1A
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176
TestEquipmentDepot.com
Содержание 2302
Страница 2: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Страница 4: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Страница 16: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Страница 22: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Страница 178: ...8 22 Status Structure Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Страница 184: ...9 6 Common Commands Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Страница 258: ...14 20 Calibration Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Страница 259: ...15 Disassembly Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Страница 265: ...16 ReplaceableParts Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Страница 296: ...16 32 Replaceable Parts Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Страница 297: ...A Specifications Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Страница 298: ...A 2 Specifications Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Страница 299: ...B ErrorandStatusMessages Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Страница 305: ...C CalibrationReference Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Страница 313: ...D CalibrationProgram Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Страница 318: ...D 6 Calibration Program Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Страница 326: ...E 8 Applications Guide Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Страница 330: ...F 4 Model 2302 Specifics Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Страница 344: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Страница 346: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Страница 347: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...