Getting Started 1-15
Table 1-3
Main MENU structure (accessed by pressing the MENU key on the Front Panel)
Menu item
Description
Ref
GPIB ADDRESS
CURRENT RANGE #1/#2
NPLC RATE #1/#2
AVER READINGS #1/#2
SAVE SETUP
RECALL SETUP
POWER ON SETUP
CALIBRATE UNIT
VOLT PROTECT #1/#2
CURR LIM MODE #1/#2
OUTPUT RELAYS
REVISION NUMBER
SERIAL NUMBER
VFD BRIGHTNESS
OUT BANDWIDTH #1/#2
OUT IMPEDANCE #1
Set primary address (0 to 30)
Select current range:
Battery channel (#1)
Model 2306/2306-VS
Model 2306-PJ
Charger channel (#2)
5A
5A
5A
5mA
500mA
5mA
AUTO
AUTO
AUTO
Set integration rate in NPLC (0.01 to 10)
Set average reading count (1 to 10)
Save present setup in memory (SAV0–SAV4)
Recall setup from memory (RST, SAV0–SAV4)
Select power-on setup (RST, SAV0–SAV4)
Calibrate unit (see calibration sections)
Set voltage protection range (0–8V) and clamp (ON/OFF) in this
manual
Select current limit mode (LIMit or TRIP)
Close (1) or open (0) relay control circuitry (except 2306-VS)
Display firmware revision levels
Display serial number of the power supply
Set VFD display’s brightness level (OFF, FULL, 3/4, 1/2, 1/4)
Set bandwidth (HIGH, LOW)
Set battery channels impedance (0–1
Ω
)
Sect. 6
Sect. 2
Sect. 2
Sect. 2
Note 1
Note 1
Note 1
Sect. 2
Note 2
Sect. 5
Note 2
Note 3
Sect. 10
Sect. 2
Sect. 2
PULSE CURRENT #1/#2
HIGH TIME
LOW TIME
AVERAGE TIME
AUTO TIME
PULSE TIMEOUT
AVERAGE READINGS
TRIGGER DELAY
TRG LEV mA RANGE
TRG LEVEL mA
TRIG LEV RANGE
Pulse-current configuration.
Set high time integration rate (in µsec.).
Set low time integration rate (in µsec.).
Set average time integration rate (in µsec.).
Set pulse integration rates automatically.
Set pulse timeout (default is 1.000 second, incremented in
1ms steps).
Set average reading count (1 to 100).
Set trigger delay in seconds (0 to 100msec).
Model 2306-PJ — Set battery channel (#1) trigger level range
on the 500mA current range (500mA, 100mA, 10mA).
Model 2306-PJ — Set pulse current trigger level in A on the
500mA current range:
Battery channel (#1)
Charger channel (#2)
mA(500mA) 0–500mA
Use TRIGGER LEVEL menu
mA(100mA) 0–100mA
item (charger channel supports
mA(10mA) 0–10mA
5A current range only)
Model 2306. 2306-VS and 2306-PJ — Set battery channel (#1)
trigger level range (5A, 1A, 100mA) on the 5A current range.
Sect. 3
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176
TestEquipmentDepot.com
Содержание 2302
Страница 2: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Страница 4: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Страница 16: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Страница 22: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Страница 178: ...8 22 Status Structure Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Страница 184: ...9 6 Common Commands Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Страница 258: ...14 20 Calibration Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Страница 259: ...15 Disassembly Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Страница 265: ...16 ReplaceableParts Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Страница 296: ...16 32 Replaceable Parts Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Страница 297: ...A Specifications Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Страница 298: ...A 2 Specifications Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Страница 299: ...B ErrorandStatusMessages Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Страница 305: ...C CalibrationReference Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Страница 313: ...D CalibrationProgram Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Страница 318: ...D 6 Calibration Program Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Страница 326: ...E 8 Applications Guide Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Страница 330: ...F 4 Model 2302 Specifics Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Страница 344: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Страница 346: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Страница 347: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...