14-14
Calibration
14.
Note and record the DMM voltage reading, then calculate the current from the voltage
reading and actual resistance value. Send that current value as the parameter for the
following command:
:CAL:PROT:STEP8 <Calculated_Current>
Example: for the Model 2306/2306-VS, with a current of 4.5mA, the command would
be:
:CAL:PROT:STEP8 4.5E-3
Example: for the Model 2306-PJ, with a current of 450mA, the command would be:
:CAL:PROT:STEP8 450E-3
Step 3: Perform Channel #2 (charger channel) calibration steps
1.
Connect the OUTPUT #2 SOURCE, SENSE, and DVM IN terminals to the digital
multimeter. Use the same general connections shown in
, but make your
connections to the OUTPUT #2 terminals. (Connect , SENSE +, and DVM
IN + to DMM INPUT HI; SOURCE -, SENSE -, and DVM IN - to DMM INPUT LO.)
2.
Send the following command to output 14V:
:CAL:PROT:STEP9 14
3.
Note and record the DMM reading, then send that reading as the parameter for the
following command:
:CAL:PROT:STEP10 <DMM_Reading>
For example, if the DMM reading is 14.012V, the command would be:
:CAL:PROT:STEP10 14.012
4.
Note and record a new DMM reading, then send that reading as the parameter for the
following command:
:CAL:PROT:STEP11 <DMM_Reading>
5.
Send the following command for DVM full-scale calibration:
:CAL:PROT:STEP12
6.
Connect the Model 2306 OUTPUT #2 SOURCE terminals to the DMM volts input and
characterized 4
Ω
resistor. Use the connections shown in
, but make your con-
nections to the OUTPUT #2 terminals instead. Be sure to observe proper polarity
( to INPUT HI; SOURCE - to INPUT LO).
7.
Make sure the digital multimeter DC volts function and auto-ranging are still selected.
8.
Send the following command to output 1.9A for 5A full-scale calibration:
:CAL:PROT:STEP13 1.9
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176
TestEquipmentDepot.com
Содержание 2302
Страница 2: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Страница 4: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Страница 16: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Страница 22: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Страница 178: ...8 22 Status Structure Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Страница 184: ...9 6 Common Commands Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Страница 258: ...14 20 Calibration Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Страница 259: ...15 Disassembly Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Страница 265: ...16 ReplaceableParts Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Страница 296: ...16 32 Replaceable Parts Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Страница 297: ...A Specifications Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Страница 298: ...A 2 Specifications Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Страница 299: ...B ErrorandStatusMessages Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Страница 305: ...C CalibrationReference Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Страница 313: ...D CalibrationProgram Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Страница 318: ...D 6 Calibration Program Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Страница 326: ...E 8 Applications Guide Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Страница 330: ...F 4 Model 2302 Specifics Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Страница 344: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Страница 346: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Страница 347: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...