Long Integration Measurements 4-19
Programming examples
The battery channel programming example applies to the Models 2306, 2306-VS, and
2306-PJ on the 5A current range. To modify the example for the 500mA current range
(Model 2306-PJ only):
1.
Change the SENS:CURR:RANG command line to select the 500mA current range.
2.
Change the trigger level commands to appropriate commands for the 500mA current
range. In the examples, the command lines requiring this modification are italicized.
The following command sequence will trigger and return one long integration measurement:
Battery channel (#1)
DISP:CHAN 1
‘ Sets display to battery channel.
SENS:CURR:RANG 5
‘ Select 5A range.
VOLT 15
‘ Set output voltage to 15V.
CURR 0.75
‘ Set current limit to 750mA.
OUTP ON
‘ Turn output on.
SENS:LINT:TLEV:ONE 0.1
‘ Set trigger level to 100mA for 1A trigger
level range.
SENS:LINT:TLEV:RANG 0.5
‘ Select 1A trigger level range.
SENS:LINT:TEDG RISING
‘ Select rising trigger edge to initiate
measurement.
SENS:LINT:TIME:AUTO
‘ Set integration time automatically for
single pulse.
SENS:FUNC “LINT”
‘ Select long integration function.
READ?
‘ Trigger and return one reading and reading
shown on display.
Charger channel (#2)
DISP:CHAN 2
‘ Sets display to charger channel.
SENS2:CURR:RANG 5
‘ Select 5A range.
SOUR2:VOLT 15
‘ Set output voltage to 15V.
SOUR2:CURR 0.75
‘ Set current limit to 750mA.
OUTP2 ON
‘ Turn output on.
SENS2:LINT:TEDge RISING
‘ Select rising trigger edge to initiate
‘ measurement.
SENS2:LINT:TLEV 0.1
‘ Set trigger level to 100mA.
SENS2:LINT:TIME:AUTO
‘ Set integration time automatically for
single pulse.
SENS2:FUNC “LINT”
‘ Select long integration function.
READ2?
‘ Trigger and return one reading and reading
shown on display
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176
TestEquipmentDepot.com
Содержание 2302
Страница 2: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Страница 4: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Страница 16: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Страница 22: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Страница 178: ...8 22 Status Structure Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Страница 184: ...9 6 Common Commands Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Страница 258: ...14 20 Calibration Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Страница 259: ...15 Disassembly Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Страница 265: ...16 ReplaceableParts Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Страница 296: ...16 32 Replaceable Parts Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Страница 297: ...A Specifications Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Страница 298: ...A 2 Specifications Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Страница 299: ...B ErrorandStatusMessages Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Страница 305: ...C CalibrationReference Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Страница 313: ...D CalibrationProgram Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Страница 318: ...D 6 Calibration Program Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Страница 326: ...E 8 Applications Guide Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Страница 330: ...F 4 Model 2302 Specifics Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Страница 344: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Страница 346: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Страница 347: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...