Section 7: SCPI command reference
Series 2280 Precision DC Power Supply Reference Manual
7-12
077085503 / March 2019
:CONFigure[n]:<function>
This command configures the instrument to a specific setup for measurements.
Type
Affected by
Where saved
Default value
Command only
Recall settings
Instrument reset
Power cycle
Save settings
See
Details
Usage
:CONFigure[n]:<function>
:CONFigure[n]:<function> <range>
:CONFigure[n]:<function> <range>,<resolution>
[n]
Channel number: For single-channel instruments, the value is
1
<function>
The measurement function to which this setting applies:
Concurrent:
CONCurrent[:DC]
Current:
CURRent[:DC]
Voltage:
VOLTage[:DC]
<range>
The range value.
Current range and concurrent range:
10 mA range:
0.01
100 mA range:
0.1
1 A range:
1
10 A range:
10
Voltage range:
Model 2280S-32-6: 32 V
Model 2280S-60-3: 60 V
<resolution>
4.5 digit resolution:
4
5.5 digit resolution:
5
6.5 digit resolution:
6
Details
This command specifies which range and resolution are used for the selected measurement function.
If you send only the command without any parameters, the present setting values for range and for
the selected measurement function are used.
Example
:CONF:CURR 0.01,5
Set the source current to the 10 mA range and the
resolution to 5.5 digits.
Also see
(on page 7-67)
:SENSe[n]:<function>:RESolution
(on page 7-73)