Section 10: Adjustment code glossary
Series 2260B Programmable Power Supplies Verification and Adjustment Manual
10-2
077104802 / November 2016
Adjustment codes
Voltage display Current display
Description
F-01
X
Power applied; Function key lights up
F-00
0000
Enter adjustment mode
C000
0
Password entered; instrument ready for adjustment
C000
1
Enter constant voltage adjustment mode
C000
2
Exit adjustment mode
C010
- - - -
Standby to execute external voltage adjustment
C011
- - - -
Output button illuminates and CV is displayed
C020
- - - -
Standby to execute external current adjustment
C021
- - - -
Output button illuminates and CC is displayed
C030
- - - -
Exit external current adjustment
C110
- - - -
Enter voltage adjustment mode
C111
- - - -
Standby adjustment offset voltage value
C112
X
Adjustment offset voltage value
C113
- - - -
Standby adjustment half-scale voltage value
C114
X
Adjustment full-scale voltage value
C115
- - - -
Standby adjustment full-scale voltage value
C116
X
Adjustment full-scale voltage value
C117
- - - -
Standby OVP adjustment
C118
ok
OVP adjustment
C120
- - - -
Enter current adjustment mode
C121
- - - -
Standby adjustment half-scale current value
C122
X
Adjustment offset current value
C123
- - - -
Standby adjustment half-scale current value
C124
X
Adjustment half-scale current value
C125
- - - -
Adjustment full-scale current value
C126
X
Set the full-scale value
C127
- - - -
Standby OCP adjustment
C128
ok
OCP adjustment
C130
0
Save mode
C130
1
Save adjustment results
C130
ok
Indicates adjustment results saved
X
X
Exit adjustment mode
The "x" in the previous table indicates that the value is not fixed.