Section 5: SCPI command reference
Series 2230G Triple-Channel Power Supplies User's Manual
5-26
2230G-900-01 Rev. A / June 2018
SOURce subsystem
The commands in the SOURce subsystem are used to control the output of the power supply.
[SOURce:]APPLy
This command sets voltage and current levels on a specified channel with a single command message.
Group
Source
Syntax
[SOURce:]APPLy {CH1|CH2|CH3}, <NRf+>, <NRf+>
Arguments
CH1|CH2|CH3
The channel to apply the settings to
<NRF+>
The voltage value to apply; valid values are
MAX|MIN|DEF|UP|DOWN
<NRF+>
The current value to apply; valid values are
MAX|MIN|DEF|UP|DOWN
Default value
CH1, 1 V, 0.1 A
Returns
Not applicable
Example
APPL CH1, 15.0, 1.0
Sets the voltage and current levels on channel 1.
See also
(on page 5-20)
(on page 5-21)
[SOURce:]CURRent[:LEVel][:IMMediate][:AMPLitude]
(on page 5-28)
[SOURce:]VOLTage[:LEVel][:IMMediate][:AMPLitude]
(on page 5-35)