5-4
Ratio and Delta
Filter, Rel, and Ranging considerations
Filter considerations
As explained in
, a unique Filter configuration can be established for each voltage
channel. However, the Filter configuration for Channel 1 is applied to both channels when Ratio
is enabled. The Filter state and configuration for Channel 2 are ignored. Channel 1 Filter has
priority because it has the most sensitive measurement range (10mV) and may therefore be
configured to provide more filtering than Channel 2.
When the FILT annunciator is on while in Ratio, the Channel 1 (DCV1) Filter settings are
applied to both input channels. When the FILT annunciator is off, filtering is not used.
When using Filter, Ratio is calculated as follows:
Ratio = Filt V1 / Filt V2
where: Filt V1 is the filtered reading for Channel 1 voltage input.
Filt V2 is the filtered reading for Channel 2 voltage input.
Keep in mind that the Filter settings are applied to the input channels, not on the result of
Ratio.
The FILT key is operational while in Ratio. Pressing FILT will either disable Filter for both
channels or enable Filter for both channels (FILT annunciator turns on). However, remember
that even though Filter can be enabled for both channels, only the Channel 1 Filter settings are
used.
NOTE
The filter configuration menu cannot be accessed while in Ratio. To make filter
configuration changes, you must first disable Ratio. This can be done by returning to
Channel 1 (press DCV1).
Rel (Relative) considerations
As explained in
, a separate Rel value can be established for each voltage channel.
When Ratio is enabled, any established Rel values are applied to the respective channels before
the calculation is performed.
Ratio is calculated as follows:
Ratio = (Filt V1 - V1 Rel) / (Filt V2 - V2 Rel)
where: Filt V1 is the filtered reading for Channel 1 voltage input.
V1 Rel is the Rel value established for Channel 1.
Filt V2 is the filtered reading for Channel 2 voltage input.
V2 Rel is the Rel value established for Channel 2.
Содержание 2182
Страница 1: ...www tek com keithley Model 2182 2182A Nanovoltmeter User s Manual 2182A 900 01 Rev B May 2017...
Страница 18: ......
Страница 22: ......
Страница 23: ...1 Getting Started Getting Started...
Страница 41: ...2 VoltageandTemperature Measurements Voltageand Temperature Measurements...
Страница 68: ...2 28 Voltage and Temperature Measurements...
Страница 69: ...3 Range Digits Rate andFilter Range Digits Rate andFilter...
Страница 82: ...3 14 Range Digits Rate and Filter...
Страница 83: ...4 Relative mX b and Percent Relative mX b andPercent...
Страница 91: ...5 RatioandDelta Ratioand Delta...
Страница 117: ...6 Buffer Buffer...
Страница 123: ...7 Triggering Triggering...
Страница 140: ...7 18 Triggering...
Страница 141: ...8 Limits Limits...
Страница 149: ...9 SteppingandScanning Steppingand Scanning...
Страница 168: ...9 20 Stepping and Scanning...
Страница 169: ...10 AnalogOutput Analog Output...
Страница 175: ...11 RemoteOperation Remote Operation...
Страница 205: ...12 CommonCommands Common Commands...
Страница 221: ...13 SCPISignalOriented Measurement Commands SCPISignalOri entedMeasure ment Commands...
Страница 225: ...14 SCPIReferenceTables SCPIRefer enceTables...
Страница 239: ...15 AdditionalSCPI Commands Additional SCPICom mands...
Страница 260: ...15 22 Additional SCPI Commands...
Страница 261: ...A Specifications Specifications...
Страница 263: ...B Statusand ErrorMessages StatusandError Messages...
Страница 268: ...B 6 Status and Error Messages...
Страница 269: ...C Measurement Considerations Measurement Consider ations...
Страница 278: ...C 10 Measurement Considerations...
Страница 279: ...D Model182Emulation Commands Model182 EmulationCom mands...
Страница 284: ...D 6 Model 182 Emulation Commands...
Страница 285: ...E Example Programs ExamplePro grams...
Страница 293: ...F IEEE 488 BusOverview IEEE 488Bus Overview...
Страница 307: ...G IEEE 488andSCPI ConformanceInformation IEEE 488and SCPIConform anceInforma tion...
Страница 310: ...G 4 IEEE 488 and SCPI Conformance Information...
Страница 311: ...H Measurement Queries Measurement Queries...
Страница 316: ...H 6 Measurement Queries...
Страница 317: ...I Delta PulseDelta and DifferentialConductance Delta Pulse Deltaand Dif ferentialCon ductance...