Relative, mX+b, and Percent (%)
4-7
Percent (%)
This math function determines percent deviation from a specified reference value. The
percent calculation is performed as follows:
Input – Reference
Percent = –––––––––––––––––
×
100%
Reference
where:
Input is the normal display reading
Reference is the user entered constant
Percent is the displayed result
To configure and control the percent calculation, perform the following steps:
1.
Press SHIFT and then % to display the present reference value:
REF: +1.000000 ^
(factory default)
2.
Key in a reference value. The
and
keys control cursor position and the
and
range keys increment and decrement the digit value. To change range, place the cursor
on the multiplier and use the
and
keys (m =
×
0.001, ^ =
×
1, K =
×
1000, and
M =
×
1,000,000). With the cursor on the polarity sign, the
and
keys toggle
polarity.
3.
Press ENTER. The MATH annunciator will turn on, and the result of the calculation will
be displayed. Note that the calculation will be applied to all measurement functions.
4.
To disable percent, again press SHIFT and then %. The MATH annunciator will turn off.
NOTES
The result of the percent calculation is positive when the input exceeds the reference
and negative when the input is less than the reference.
The result of the percent calculation may be displayed in exponential notation. For
example a displayed reading of + 2.500E+03% is equivalent to 2500% (2.5K %).
Содержание 2182
Страница 1: ...www tek com keithley Model 2182 2182A Nanovoltmeter User s Manual 2182A 900 01 Rev B May 2017...
Страница 18: ......
Страница 22: ......
Страница 23: ...1 Getting Started Getting Started...
Страница 41: ...2 VoltageandTemperature Measurements Voltageand Temperature Measurements...
Страница 68: ...2 28 Voltage and Temperature Measurements...
Страница 69: ...3 Range Digits Rate andFilter Range Digits Rate andFilter...
Страница 82: ...3 14 Range Digits Rate and Filter...
Страница 83: ...4 Relative mX b and Percent Relative mX b andPercent...
Страница 91: ...5 RatioandDelta Ratioand Delta...
Страница 117: ...6 Buffer Buffer...
Страница 123: ...7 Triggering Triggering...
Страница 140: ...7 18 Triggering...
Страница 141: ...8 Limits Limits...
Страница 149: ...9 SteppingandScanning Steppingand Scanning...
Страница 168: ...9 20 Stepping and Scanning...
Страница 169: ...10 AnalogOutput Analog Output...
Страница 175: ...11 RemoteOperation Remote Operation...
Страница 205: ...12 CommonCommands Common Commands...
Страница 221: ...13 SCPISignalOriented Measurement Commands SCPISignalOri entedMeasure ment Commands...
Страница 225: ...14 SCPIReferenceTables SCPIRefer enceTables...
Страница 239: ...15 AdditionalSCPI Commands Additional SCPICom mands...
Страница 260: ...15 22 Additional SCPI Commands...
Страница 261: ...A Specifications Specifications...
Страница 263: ...B Statusand ErrorMessages StatusandError Messages...
Страница 268: ...B 6 Status and Error Messages...
Страница 269: ...C Measurement Considerations Measurement Consider ations...
Страница 278: ...C 10 Measurement Considerations...
Страница 279: ...D Model182Emulation Commands Model182 EmulationCom mands...
Страница 284: ...D 6 Model 182 Emulation Commands...
Страница 285: ...E Example Programs ExamplePro grams...
Страница 293: ...F IEEE 488 BusOverview IEEE 488Bus Overview...
Страница 307: ...G IEEE 488andSCPI ConformanceInformation IEEE 488and SCPIConform anceInforma tion...
Страница 310: ...G 4 IEEE 488 and SCPI Conformance Information...
Страница 311: ...H Measurement Queries Measurement Queries...
Страница 316: ...H 6 Measurement Queries...
Страница 317: ...I Delta PulseDelta and DifferentialConductance Delta Pulse Deltaand Dif ferentialCon ductance...