Range, Digits, Rate, and Filter
3-7
NOTE
For remote operation, the integration time can be set from 0.01 PLC to 60 PLC
(50 PLC for 50Hz line power). Integration time can instead be set as an aperture time
from 166.67µsec (200µsec for 50Hz) to 1 second.
Perform the following steps to set the integration rate:
1.
Select the desired function.
2.
Press the RATE key until the desired number of power line cycles (PLC) is displayed.
The appropriate annunciator will turn on (FAST, MED, or SLOW).
NOTE
Pressing the front panel RATE key will enable Autozero if it was off. For remote
programming, the rate commands have no effect on the state of Autozero. For details,
see “
.
SCPI programming - rate
Programming example - rate
The following program fragment sets the voltage reading rate to 2 PLC and the temperature
reading rate to 5 PLC:
CALL SEND(7,“:sens:volt:nplc 2”,status%)
‘Set volts for 2 PLC.
CALL SEND(7,“:sens:temp:nplc 5”,status%)
‘Set temp for 5 PLC.
Table 3-3
SCPI commands - rate
Commands
Description
Default
:SENSe
SENSe Subsystem:
:VOLTage
DCV1 and DCV2:
:NPLCycles <n>
:APERture <n>
Specify integration rate in PLCs: 0.01 to 60 (60Hz)
0.01 to 50 (50Hz)
Specify integration rate in seconds: 166.67µsec to 1 sec (60Hz)
200µsec to 1 sec (50Hz)
5
83.33msec
:TEMPerature
TEMP1 and TEMP2:
:NPLCycles <n>
:APERture <n>
Specify integration rate in PLCs: 0.01 to 60 (60Hz)
0.01 to 50 (50Hz)
Specify integration rate in seconds: 166.67µsec to 1 sec (60Hz)
200µsec to 1 sec (50Hz)
5
83.33msec
Содержание 2182
Страница 1: ...www tek com keithley Model 2182 2182A Nanovoltmeter User s Manual 2182A 900 01 Rev B May 2017...
Страница 18: ......
Страница 22: ......
Страница 23: ...1 Getting Started Getting Started...
Страница 41: ...2 VoltageandTemperature Measurements Voltageand Temperature Measurements...
Страница 68: ...2 28 Voltage and Temperature Measurements...
Страница 69: ...3 Range Digits Rate andFilter Range Digits Rate andFilter...
Страница 82: ...3 14 Range Digits Rate and Filter...
Страница 83: ...4 Relative mX b and Percent Relative mX b andPercent...
Страница 91: ...5 RatioandDelta Ratioand Delta...
Страница 117: ...6 Buffer Buffer...
Страница 123: ...7 Triggering Triggering...
Страница 140: ...7 18 Triggering...
Страница 141: ...8 Limits Limits...
Страница 149: ...9 SteppingandScanning Steppingand Scanning...
Страница 168: ...9 20 Stepping and Scanning...
Страница 169: ...10 AnalogOutput Analog Output...
Страница 175: ...11 RemoteOperation Remote Operation...
Страница 205: ...12 CommonCommands Common Commands...
Страница 221: ...13 SCPISignalOriented Measurement Commands SCPISignalOri entedMeasure ment Commands...
Страница 225: ...14 SCPIReferenceTables SCPIRefer enceTables...
Страница 239: ...15 AdditionalSCPI Commands Additional SCPICom mands...
Страница 260: ...15 22 Additional SCPI Commands...
Страница 261: ...A Specifications Specifications...
Страница 263: ...B Statusand ErrorMessages StatusandError Messages...
Страница 268: ...B 6 Status and Error Messages...
Страница 269: ...C Measurement Considerations Measurement Consider ations...
Страница 278: ...C 10 Measurement Considerations...
Страница 279: ...D Model182Emulation Commands Model182 EmulationCom mands...
Страница 284: ...D 6 Model 182 Emulation Commands...
Страница 285: ...E Example Programs ExamplePro grams...
Страница 293: ...F IEEE 488 BusOverview IEEE 488Bus Overview...
Страница 307: ...G IEEE 488andSCPI ConformanceInformation IEEE 488and SCPIConform anceInforma tion...
Страница 310: ...G 4 IEEE 488 and SCPI Conformance Information...
Страница 311: ...H Measurement Queries Measurement Queries...
Страница 316: ...H 6 Measurement Queries...
Страница 317: ...I Delta PulseDelta and DifferentialConductance Delta Pulse Deltaand Dif ferentialCon ductance...