Additional SCPI Commands
15-9
Questionable Event Register:
Bits B0 through B3
– Not used.
Bit B4, Temperature Summary (Temp)
– Set bit indicates that an invalid reference junction
measurement has occurred for thermocouple temperature measurements.
Bits B5 through B7
– Not used.
Bit B8, Calibration Summary (Cal)
– Set bit indicates that an invalid calibration constant
was detected during the power-up sequence. The instrument will instead use a default
calibration constant. This error will clear after successful calibration of the instrument.
Bit B9, ACAL Summary (ACAL)
– Set bit indicates that an invalid ACAL was performed.
This error will clear after a successful ACAL is performed.
Bits B10 through B15
– Not used.
NOTE
Whenever a questionable event occurs, the ERR annunciator will turn on. The
annunciator will turn off when the questionable event clears.
Figure 15-5
Questionable event register
Cal
B15 - B10
B9
B8 B7 -B5 B4
Bit
Position
Event
Decimal
Weighting
Value
Value : 1 = Operation Event Set
0 = Operation Event Cleared
Events : ACAL = ACAL Summary
Cal = Calibration Summary
Temp = Temperature Summary
Temp
(2 )
4
16
0/1
B3 - B0
ACAL
(2 )
9
512
(2 )
8
256
0/1
0/1
Содержание 2182
Страница 1: ...www tek com keithley Model 2182 2182A Nanovoltmeter User s Manual 2182A 900 01 Rev B May 2017...
Страница 18: ......
Страница 22: ......
Страница 23: ...1 Getting Started Getting Started...
Страница 41: ...2 VoltageandTemperature Measurements Voltageand Temperature Measurements...
Страница 68: ...2 28 Voltage and Temperature Measurements...
Страница 69: ...3 Range Digits Rate andFilter Range Digits Rate andFilter...
Страница 82: ...3 14 Range Digits Rate and Filter...
Страница 83: ...4 Relative mX b and Percent Relative mX b andPercent...
Страница 91: ...5 RatioandDelta Ratioand Delta...
Страница 117: ...6 Buffer Buffer...
Страница 123: ...7 Triggering Triggering...
Страница 140: ...7 18 Triggering...
Страница 141: ...8 Limits Limits...
Страница 149: ...9 SteppingandScanning Steppingand Scanning...
Страница 168: ...9 20 Stepping and Scanning...
Страница 169: ...10 AnalogOutput Analog Output...
Страница 175: ...11 RemoteOperation Remote Operation...
Страница 205: ...12 CommonCommands Common Commands...
Страница 221: ...13 SCPISignalOriented Measurement Commands SCPISignalOri entedMeasure ment Commands...
Страница 225: ...14 SCPIReferenceTables SCPIRefer enceTables...
Страница 239: ...15 AdditionalSCPI Commands Additional SCPICom mands...
Страница 260: ...15 22 Additional SCPI Commands...
Страница 261: ...A Specifications Specifications...
Страница 263: ...B Statusand ErrorMessages StatusandError Messages...
Страница 268: ...B 6 Status and Error Messages...
Страница 269: ...C Measurement Considerations Measurement Consider ations...
Страница 278: ...C 10 Measurement Considerations...
Страница 279: ...D Model182Emulation Commands Model182 EmulationCom mands...
Страница 284: ...D 6 Model 182 Emulation Commands...
Страница 285: ...E Example Programs ExamplePro grams...
Страница 293: ...F IEEE 488 BusOverview IEEE 488Bus Overview...
Страница 307: ...G IEEE 488andSCPI ConformanceInformation IEEE 488and SCPIConform anceInforma tion...
Страница 310: ...G 4 IEEE 488 and SCPI Conformance Information...
Страница 311: ...H Measurement Queries Measurement Queries...
Страница 316: ...H 6 Measurement Queries...
Страница 317: ...I Delta PulseDelta and DifferentialConductance Delta Pulse Deltaand Dif ferentialCon ductance...