15-8
Additional SCPI Commands
Measurement Event Register:
Bit B0, Reading Overflow (ROF)
– Set bit indicates that the reading exceeds the measure-
ment range of the instrument.
Bit B1, Low Limit1 (LL1)
– Set bit indicates that the reading is less than the Low Limit 1
setting.
Bit B2, High Limit1 (HL1)
– Set bit indicates that the reading is greater than the High Limit
1 setting.
Bit B3, Low Limit 2 (LL2)
– Set bit indicates that the reading is less than the Low Limit 2
setting.
Bit B4, High Limit 2 (HL2)
– Set bit indicates that the reading is greater than the High Limit
2 setting.
Bit B5, Reading Available (RAV)
– Set bit indicates that a reading was taken and processed.
Bit B6
– Not used.
Bit B7, Buffer Available (BAV)
– Set bit indicates that there are at least two readings in the
trace buffer.
Bit B8, Buffer Half Full (BHF)
– Set bit indicates that the trace buffer is half full.
Bit B9, Buffer Full (BFL)
– Set bit indicates that the trace buffer is full.
Bits B10 through B15
– Not used.
Figure 15-4
Measurement event register
RAV
B15 - B10
B9
B8
B7
B6
B5
B4
B3
B2
B1
B0
(2 )
9
(2 )
8
(2 )
7
(2 )
5
(2 )
2
(2 )
1
(2 )
0
32
0/1
Bit Position
Event
Decimal Weighting
Value
Value : 1 = Measurement Event Set
0 = Measurement Event Cleared
LL1
2
0/1
Events : BFL = Buffer Full
BHF = Buffer Half Full
BAV = Buffer Available
RAV = Reading Available
HL1 = High Limit 1
HL2 = High Limit 2
LL1 = Low Limit 1
LL2 = Low Limit 2
ROF = Reading Overflow
BFL
HL1
ROF
512
4
1
0/1
0/1
0/1
BHF BAV
256
128
0/1
0/1
LL2
HL2
0/1
0/1
(2 )
3
8
(2 )
4
16
Содержание 2182
Страница 1: ...www tek com keithley Model 2182 2182A Nanovoltmeter User s Manual 2182A 900 01 Rev B May 2017...
Страница 18: ......
Страница 22: ......
Страница 23: ...1 Getting Started Getting Started...
Страница 41: ...2 VoltageandTemperature Measurements Voltageand Temperature Measurements...
Страница 68: ...2 28 Voltage and Temperature Measurements...
Страница 69: ...3 Range Digits Rate andFilter Range Digits Rate andFilter...
Страница 82: ...3 14 Range Digits Rate and Filter...
Страница 83: ...4 Relative mX b and Percent Relative mX b andPercent...
Страница 91: ...5 RatioandDelta Ratioand Delta...
Страница 117: ...6 Buffer Buffer...
Страница 123: ...7 Triggering Triggering...
Страница 140: ...7 18 Triggering...
Страница 141: ...8 Limits Limits...
Страница 149: ...9 SteppingandScanning Steppingand Scanning...
Страница 168: ...9 20 Stepping and Scanning...
Страница 169: ...10 AnalogOutput Analog Output...
Страница 175: ...11 RemoteOperation Remote Operation...
Страница 205: ...12 CommonCommands Common Commands...
Страница 221: ...13 SCPISignalOriented Measurement Commands SCPISignalOri entedMeasure ment Commands...
Страница 225: ...14 SCPIReferenceTables SCPIRefer enceTables...
Страница 239: ...15 AdditionalSCPI Commands Additional SCPICom mands...
Страница 260: ...15 22 Additional SCPI Commands...
Страница 261: ...A Specifications Specifications...
Страница 263: ...B Statusand ErrorMessages StatusandError Messages...
Страница 268: ...B 6 Status and Error Messages...
Страница 269: ...C Measurement Considerations Measurement Consider ations...
Страница 278: ...C 10 Measurement Considerations...
Страница 279: ...D Model182Emulation Commands Model182 EmulationCom mands...
Страница 284: ...D 6 Model 182 Emulation Commands...
Страница 285: ...E Example Programs ExamplePro grams...
Страница 293: ...F IEEE 488 BusOverview IEEE 488Bus Overview...
Страница 307: ...G IEEE 488andSCPI ConformanceInformation IEEE 488and SCPIConform anceInforma tion...
Страница 310: ...G 4 IEEE 488 and SCPI Conformance Information...
Страница 311: ...H Measurement Queries Measurement Queries...
Страница 316: ...H 6 Measurement Queries...
Страница 317: ...I Delta PulseDelta and DifferentialConductance Delta Pulse Deltaand Dif ferentialCon ductance...