Analog Output
10-5
Operation
Analog output connections
The analog output is accessed from the rear panel BNC connector that is labeled “ANALOG
OUTPUT.” This connector requires a cable that is terminated with a standard male BNC connector.
Output resistance
— The output resistance of Analog Output is 1k
Ω
±5%. To minimize the
effects of loading, the input resistance of the device connected to Analog Output should be as
high as possible. For example, with a device with an input resistance of 10M
Ω
, the error due to
loading will be approximately 0.01%.
Configure and control analog output
1.
Press SHIFT and then AOUT to display the Gain (M) factor:
M +1.0000000^ (factory default)
2.
Gain can be set from -100e6 to 100e6. The
and
keys control cursor position and
the
and
range keys increment and decrement the digit value. To change range,
place the cursor on the multiplier and use the
and
keys (m =
×
0.001, ^ =
×
1,
K =
×
1000, and M =
×
1,000,000). With the cursor on the polarity sign, the
and
keys toggle polarity.
3.
Press ENTER to enter the Gain value and display the Offset value:
B: +00.000000 (factory default)
4.
Key in the Offset value.
5.
Press ENTER to enter the Offset value and enable Analog Output. The instrument
returns to the normal display state.
6.
To disable Analog Output, press SHIFT and then AOUT.
Analog output rel
With analog output enabled, Analog Output Rel is used to automatically reference the analog
output voltage to zero. When Analog Output Rel is turned ON, the present analog output voltage
is used as the Rel value. This sets the analog output voltage to zero. Subsequent analog output
readings will be the difference between the actual analog output and the Rel value.
To enable Analog Output Rel, press SHIFT and then OUTPUT. The message “AOUT REL
ON” will be displayed briefly to indicate that it is enabled. To disable Analog Output REL, press
SHIFT and then OUTPUT a second time. The message “AOUT REL OFF” will be displayed
briefly.
NOTE
A new Rel value for analog output can be established at any time by first disabling
Analog Output Rel and then re-enabling it.
Содержание 2182
Страница 1: ...www tek com keithley Model 2182 2182A Nanovoltmeter User s Manual 2182A 900 01 Rev B May 2017...
Страница 18: ......
Страница 22: ......
Страница 23: ...1 Getting Started Getting Started...
Страница 41: ...2 VoltageandTemperature Measurements Voltageand Temperature Measurements...
Страница 68: ...2 28 Voltage and Temperature Measurements...
Страница 69: ...3 Range Digits Rate andFilter Range Digits Rate andFilter...
Страница 82: ...3 14 Range Digits Rate and Filter...
Страница 83: ...4 Relative mX b and Percent Relative mX b andPercent...
Страница 91: ...5 RatioandDelta Ratioand Delta...
Страница 117: ...6 Buffer Buffer...
Страница 123: ...7 Triggering Triggering...
Страница 140: ...7 18 Triggering...
Страница 141: ...8 Limits Limits...
Страница 149: ...9 SteppingandScanning Steppingand Scanning...
Страница 168: ...9 20 Stepping and Scanning...
Страница 169: ...10 AnalogOutput Analog Output...
Страница 175: ...11 RemoteOperation Remote Operation...
Страница 205: ...12 CommonCommands Common Commands...
Страница 221: ...13 SCPISignalOriented Measurement Commands SCPISignalOri entedMeasure ment Commands...
Страница 225: ...14 SCPIReferenceTables SCPIRefer enceTables...
Страница 239: ...15 AdditionalSCPI Commands Additional SCPICom mands...
Страница 260: ...15 22 Additional SCPI Commands...
Страница 261: ...A Specifications Specifications...
Страница 263: ...B Statusand ErrorMessages StatusandError Messages...
Страница 268: ...B 6 Status and Error Messages...
Страница 269: ...C Measurement Considerations Measurement Consider ations...
Страница 278: ...C 10 Measurement Considerations...
Страница 279: ...D Model182Emulation Commands Model182 EmulationCom mands...
Страница 284: ...D 6 Model 182 Emulation Commands...
Страница 285: ...E Example Programs ExamplePro grams...
Страница 293: ...F IEEE 488 BusOverview IEEE 488Bus Overview...
Страница 307: ...G IEEE 488andSCPI ConformanceInformation IEEE 488and SCPIConform anceInforma tion...
Страница 310: ...G 4 IEEE 488 and SCPI Conformance Information...
Страница 311: ...H Measurement Queries Measurement Queries...
Страница 316: ...H 6 Measurement Queries...
Страница 317: ...I Delta PulseDelta and DifferentialConductance Delta Pulse Deltaand Dif ferentialCon ductance...