Stepping and Scanning
9-7
Step/Scan configuration
Internal Stepping/Scanning
The settings for internal stepping and scanning are explained as follows:
Timer
— The maximum timer interval is 99H:99M:99.999S (Hour:Minute:Second format).
Channel 1 Count
— This specifies the number of measurements to be performed while on
Channel 1. Keep in mind that for each step/scan cycle, only one measurement is performed on
Channel 2. Channel 1 Count can be set from 1 to 1023.
Reading Count
— This indicates the total number of measurements that will to be performed
for each step/scan cycle. For example, if Channel 1 Count is set to 3, the Reading Count will
initially set to 4 (the extra reading is for Channel 2). For each additional step/scan cycle, simply
add 4 to the Reading Count. Therefore, to perform three step/scan cycles, set the Reading Count
to 12. Reading Count can be set from 2 to 1024.
NOTES
• If you change the Reading Count, it must be a multiple of the initial count val-
ue. For example, if the initial Reading Count is 3, you can change it to 6 (or
9, or 12, etc.). If you enter a non-multiple value, the instrument will select the
next lower value that is a multiple.
• Reading Count can be set to a value >1024 or INFinite, but only the first 1024
readings will be stored in the buffer.
Perform the following steps to configure internal stepping or scanning:
1.
Press SHIFT and then CONFIG. Use the
key to display the present SCANNING type
(INTernal or EXTernal).
2.
Press the
or
key to display INT, and press ENTER.
3.
The present state of the timer is displayed (OFF or ON). Press
or
to display the
desired timer state, and press ENTER.
4.
If you turned the timer on, the timer interval will be displayed. Use the edit keys (
,
,
, and
) to set the timer interval, and press ENTER.
5.
The present Channel 1 Count (CH1 CNT) is displayed. Use the edit keys to set the
number of measurements for Channel 1, and press ENTER.
6.
The Present Reading Count (RNG CNT) is displayed (it will be CH1 CNT + 1). If you
wish to increase the Reading Count, use the edit keys to display the value, and press
ENTER to return to the normal display state.
Содержание 2182
Страница 1: ...www tek com keithley Model 2182 2182A Nanovoltmeter User s Manual 2182A 900 01 Rev B May 2017...
Страница 18: ......
Страница 22: ......
Страница 23: ...1 Getting Started Getting Started...
Страница 41: ...2 VoltageandTemperature Measurements Voltageand Temperature Measurements...
Страница 68: ...2 28 Voltage and Temperature Measurements...
Страница 69: ...3 Range Digits Rate andFilter Range Digits Rate andFilter...
Страница 82: ...3 14 Range Digits Rate and Filter...
Страница 83: ...4 Relative mX b and Percent Relative mX b andPercent...
Страница 91: ...5 RatioandDelta Ratioand Delta...
Страница 117: ...6 Buffer Buffer...
Страница 123: ...7 Triggering Triggering...
Страница 140: ...7 18 Triggering...
Страница 141: ...8 Limits Limits...
Страница 149: ...9 SteppingandScanning Steppingand Scanning...
Страница 168: ...9 20 Stepping and Scanning...
Страница 169: ...10 AnalogOutput Analog Output...
Страница 175: ...11 RemoteOperation Remote Operation...
Страница 205: ...12 CommonCommands Common Commands...
Страница 221: ...13 SCPISignalOriented Measurement Commands SCPISignalOri entedMeasure ment Commands...
Страница 225: ...14 SCPIReferenceTables SCPIRefer enceTables...
Страница 239: ...15 AdditionalSCPI Commands Additional SCPICom mands...
Страница 260: ...15 22 Additional SCPI Commands...
Страница 261: ...A Specifications Specifications...
Страница 263: ...B Statusand ErrorMessages StatusandError Messages...
Страница 268: ...B 6 Status and Error Messages...
Страница 269: ...C Measurement Considerations Measurement Consider ations...
Страница 278: ...C 10 Measurement Considerations...
Страница 279: ...D Model182Emulation Commands Model182 EmulationCom mands...
Страница 284: ...D 6 Model 182 Emulation Commands...
Страница 285: ...E Example Programs ExamplePro grams...
Страница 293: ...F IEEE 488 BusOverview IEEE 488Bus Overview...
Страница 307: ...G IEEE 488andSCPI ConformanceInformation IEEE 488and SCPIConform anceInforma tion...
Страница 310: ...G 4 IEEE 488 and SCPI Conformance Information...
Страница 311: ...H Measurement Queries Measurement Queries...
Страница 316: ...H 6 Measurement Queries...
Страница 317: ...I Delta PulseDelta and DifferentialConductance Delta Pulse Deltaand Dif ferentialCon ductance...