9-4
Stepping and Scanning
Front panel trigger models
The front panel trigger models for stepping and scanning are shown in
. These are expansions of the basic front panel trigger model that is presented and
explained in
). The following discussions focus on the stepping and
scanning operations. Be sure to refer to
for additional information on the various
components of trigger models.
Internal scanning
shows the front panel trigger model for internal scanning. The components of the
trigger model are explained as follows:
•
Control source
:
•
Immediate
— With immediate triggering, event detection occurs immediately
allowing operation to drop down to the next trigger model block (Delay).
•
Timer
— The timer is used to control the time interval between internal scans. It
has no effect on the time interval between each measurement in a scan cycle. When
SCAN is pressed, the timer starts and event detection occurs immediately allowing
operation to drop down to Delay.
If configured for an additional scan, operation will later loop back to this control
source, where it will wait until the timer interval expires. If the timer interval is
already expired, event detection will be satisfied immediately.
•
External trigger
— After the internal scan is configured, pressing the EX TRIG
key places the instrument in the external trigger mode. When the SCAN key is
pressed, the internal scan is enabled. However, it doesn’t start until an external trig-
ger is received or the TRIG key is pressed. After the trigger occurs, operation drops
down to the Delay block.
•
Delay
— If a delay (auto or manual) is being used, operation will hold up at this block
until the delay period expires.
•
Device action
— Measurements are performed at this block. The first measurement is
performed on Channel 2 of the Model 2182. Subsequent measurements are performed
on Channel 1. Note that every reading is automatically stored in the buffer.
•
Sample counter
— For each scan cycle, Channel 2 is measured once and Channel 1 is
measured a specified number of times. Therefore, the sample count is the sum of those
two values (CH1 COUNT +1). Operation will continue to loop back to Delay and Device
Action until all the sample readings are taken.
•
Output trigger
— After the last Channel 1 measurement is performed, an output trigger
is applied to the rear panel Trigger Link connector.
•
Trigger counter
— The value of this counter determines how many scan cycles will be
performed. If configured for another scan, operation will loop back to the control source
for another pass through the trigger model.
Содержание 2182
Страница 1: ...www tek com keithley Model 2182 2182A Nanovoltmeter User s Manual 2182A 900 01 Rev B May 2017...
Страница 18: ......
Страница 22: ......
Страница 23: ...1 Getting Started Getting Started...
Страница 41: ...2 VoltageandTemperature Measurements Voltageand Temperature Measurements...
Страница 68: ...2 28 Voltage and Temperature Measurements...
Страница 69: ...3 Range Digits Rate andFilter Range Digits Rate andFilter...
Страница 82: ...3 14 Range Digits Rate and Filter...
Страница 83: ...4 Relative mX b and Percent Relative mX b andPercent...
Страница 91: ...5 RatioandDelta Ratioand Delta...
Страница 117: ...6 Buffer Buffer...
Страница 123: ...7 Triggering Triggering...
Страница 140: ...7 18 Triggering...
Страница 141: ...8 Limits Limits...
Страница 149: ...9 SteppingandScanning Steppingand Scanning...
Страница 168: ...9 20 Stepping and Scanning...
Страница 169: ...10 AnalogOutput Analog Output...
Страница 175: ...11 RemoteOperation Remote Operation...
Страница 205: ...12 CommonCommands Common Commands...
Страница 221: ...13 SCPISignalOriented Measurement Commands SCPISignalOri entedMeasure ment Commands...
Страница 225: ...14 SCPIReferenceTables SCPIRefer enceTables...
Страница 239: ...15 AdditionalSCPI Commands Additional SCPICom mands...
Страница 260: ...15 22 Additional SCPI Commands...
Страница 261: ...A Specifications Specifications...
Страница 263: ...B Statusand ErrorMessages StatusandError Messages...
Страница 268: ...B 6 Status and Error Messages...
Страница 269: ...C Measurement Considerations Measurement Consider ations...
Страница 278: ...C 10 Measurement Considerations...
Страница 279: ...D Model182Emulation Commands Model182 EmulationCom mands...
Страница 284: ...D 6 Model 182 Emulation Commands...
Страница 285: ...E Example Programs ExamplePro grams...
Страница 293: ...F IEEE 488 BusOverview IEEE 488Bus Overview...
Страница 307: ...G IEEE 488andSCPI ConformanceInformation IEEE 488and SCPIConform anceInforma tion...
Страница 310: ...G 4 IEEE 488 and SCPI Conformance Information...
Страница 311: ...H Measurement Queries Measurement Queries...
Страница 316: ...H 6 Measurement Queries...
Страница 317: ...I Delta PulseDelta and DifferentialConductance Delta Pulse Deltaand Dif ferentialCon ductance...