Buffer operations
— Explains how to store and recall readings including buffer
statistics (minimum, maximum, peak-to-peak, average, and standard deviation).
•
SCPI programming
— Covers the SCPI commands used to control buffer operations.
Buffer operations
The Model 2182 has a buffer to store from two to 1024 readings and units. It also stores the
channel number for step/scan readings and overflow readings. In addition, recalled data includes
statistical information (minimum, maximum, peak-to-peak, average, and standard deviation).
NOTE
Statistics are not calculated when an overflow reading has been stored in the buffer.
The buffer fills with the specified number of readings and stops. Readings are placed in the
buffer after any math operations are performed. Math operations include Relative, Ratio or
Delta, and mX+b or Percent.
Buffered data is overwritten each time the storage operation is selected. The data is volatile;
it is not saved through a power cycle.
NOTE
Measurements performed during stepping or scanning are automatically stored in the
buffer. There is no need to configure and enable the buffer. Stepping and scanning is
covered in
.
NOTE
When changing the interface (GPIB to RS-232, or vice versa), all data in the buffer clears.
Store
Perform the following steps to store readings:
1.
Set up the instrument for the desired configuration.
2.
Press the STORE key.
3.
Use the cursor keys (
and
), and the RANGE
and
keys to set the number of
readings to store (2 to 1024).
4.
Press ENTER to enable the buffer. If in the immediate trigger mode, the storage process
will start immediately. If in the external trigger mode, each input trigger (or press of
TRIG key) will store a reading.
NOTE
The asterisk (*) annunciator turns on to indicate that the data storage operation is
enabled. It will turn off when the storage process is finished (buffer full).
6-2
Buffer
Содержание 2182
Страница 1: ...www tek com keithley Model 2182 2182A Nanovoltmeter User s Manual 2182A 900 01 Rev B May 2017...
Страница 18: ......
Страница 22: ......
Страница 23: ...1 Getting Started Getting Started...
Страница 41: ...2 VoltageandTemperature Measurements Voltageand Temperature Measurements...
Страница 68: ...2 28 Voltage and Temperature Measurements...
Страница 69: ...3 Range Digits Rate andFilter Range Digits Rate andFilter...
Страница 82: ...3 14 Range Digits Rate and Filter...
Страница 83: ...4 Relative mX b and Percent Relative mX b andPercent...
Страница 91: ...5 RatioandDelta Ratioand Delta...
Страница 117: ...6 Buffer Buffer...
Страница 123: ...7 Triggering Triggering...
Страница 140: ...7 18 Triggering...
Страница 141: ...8 Limits Limits...
Страница 149: ...9 SteppingandScanning Steppingand Scanning...
Страница 168: ...9 20 Stepping and Scanning...
Страница 169: ...10 AnalogOutput Analog Output...
Страница 175: ...11 RemoteOperation Remote Operation...
Страница 205: ...12 CommonCommands Common Commands...
Страница 221: ...13 SCPISignalOriented Measurement Commands SCPISignalOri entedMeasure ment Commands...
Страница 225: ...14 SCPIReferenceTables SCPIRefer enceTables...
Страница 239: ...15 AdditionalSCPI Commands Additional SCPICom mands...
Страница 260: ...15 22 Additional SCPI Commands...
Страница 261: ...A Specifications Specifications...
Страница 263: ...B Statusand ErrorMessages StatusandError Messages...
Страница 268: ...B 6 Status and Error Messages...
Страница 269: ...C Measurement Considerations Measurement Consider ations...
Страница 278: ...C 10 Measurement Considerations...
Страница 279: ...D Model182Emulation Commands Model182 EmulationCom mands...
Страница 284: ...D 6 Model 182 Emulation Commands...
Страница 285: ...E Example Programs ExamplePro grams...
Страница 293: ...F IEEE 488 BusOverview IEEE 488Bus Overview...
Страница 307: ...G IEEE 488andSCPI ConformanceInformation IEEE 488and SCPIConform anceInforma tion...
Страница 310: ...G 4 IEEE 488 and SCPI Conformance Information...
Страница 311: ...H Measurement Queries Measurement Queries...
Страница 316: ...H 6 Measurement Queries...
Страница 317: ...I Delta PulseDelta and DifferentialConductance Delta Pulse Deltaand Dif ferentialCon ductance...