Default calibration constants and tolerances
Default calibration constants and tolerances for the Model 2182 are listed in
. The
tolerances for user-entered values are expressed as a percentage (10%). For example, for the
+10V calibration, the calibration signal can be +9V to +11V. The rest of the tolerances are
expressed as full-scale values. Most are 10% of the default value.
The tolerance for the ACAL temperature constant (ACalTemp) is ±30°C of the default (which
is 30°C). Therefore, valid ACAL can be performed for temperatures from 0 to 60°C.