Model 2110 5½ Digit Multimeter Reference Manual
Section 3: Basic DMM operation
2110-901-01 Rev. C/August 2013
3-33
Sample count
The sample count is the number of measurements (up to 2000) to perform each time the instrument
is triggered.
The sample count is stored in volatile memory. It defaults to one (1) when instrument power is turned
off and then turned on again
Setting sample count from the front panel
Perform the following steps to set the sample count:
1. Press the
SHIFT
and
CONFIG
(
MENU
) keys to display the MENU options.
2.
Use the ◄ and ► keys to display
TRIG SYS
and press
ENTER
.
3.
Use the ◄ and ► keys to display
SAMPLE CNT
and press
ENTER.
4.
Use the ◄, ►, ▲, and ▼ keys to specify the sample count (1 to 2000) and press
ENTER.
Setting sample count using remote operation
Commands that use the
<n>
parameter can use the
MINimum
and
MAXimum
name parameters.
MIN
selects the minimum value and
MAX
selects the maximum value.
Use the following command to set the sample count:
SAMPle:COUNt <n>
<n>
= 1 to 2000
Sample count
Trigger count
For remote operation, the trigger count (up to 2000) specifies how many times the sample count
measurements can be performed. Trigger count is ignored for front-panel operation.
The trigger count is stored in volatile memory. It defaults to one (1) when instrument power is turned
off and then turned on again.
Setting trigger count using remote operation
Commands that use the
<n>
parameter can use the
MINimum
and
MAXimum
name parameters.
MIN
selects the minimum value and
MAX
selects the maximum value.
You can change the trigger count using the front panel, but the setting only applies during remote
operation.
Use the following command to set the trigger count:
TRIGger:COUNt <n>
<n>
= 1 to 2000
Trigger count