
Section 5: Remote commands
Model 2110 5½ Digit Multimeter Reference Manual
5-26
2110-901-01 Rev. C/August 2013
TRIGger subsystem summary
Command
Description
TRIGger
:SOURce <name>
Specify a trigger source. The instrument will accept a software
(BUS) trigger, an immediate internal trigger (IMMediate), or a
hardware trigger from the rear-panel (EXT TRIG) terminal.
:SOURce?
Query the trigger source.
:COUNt <n> MIN|MAX|DEF|INF
Specify the number of triggers the instrument will accept
before returning to the “idle” state. (1 to 50,000, MIN, MAX, or
INF). The INFinite parameter instructs the instrument to take
triggers continuously.
:COUNt?
Query measure count. Returns the number of triggers the
instrument will accept before returning to the “idle” state.
:DELay <n>|MIN|MAX|DEF
Set delay time (0 to 3600 sec., MIN or MAX.
:DELay? [MIN|MAX|DEF]
Query the trigger delay time
:AUTO <b>
Enable or disable auto delay. Where: 0 = off and 1 = on.
:AUTO?
Query state of delay. Returns 0 (off) or 1 (on).
:SAMPle
Path to trigger sample.
:COUNt <n>|MIN|MAX|DEF
Set the number of readings (samples) the instrument takes per
trigger (1 to 50,000, MIN or MAX)
:COUNt? [MIN|MAX|DEF]
Query the sample count.