Calibration Command Reference
3-6
3.3.8 :DC
(:CALibration:PROtected:DC)
The :DC commands perform comprehensive (user) calibration. Table 3-2 summarizes these
comprehensive calibration commands along with parameter limits.
:ZERO
(:CALibration:PROTected:DC:ZERO)
Purpose
To perform short-circuit comprehensive calibration.
Format
:cal:prot:dc:zero
Parameter
none
Description
:ZERO performs the short-circuit calibration step in the comprehensive calibration procedure.
A low-thermal short (Model 8610) must be connected to the input jacks before sending this
command.
Example
:CAL:PROT:DC:ZERO
Perform zero calibration.
Table 3-2
Comprehensive calibration commands
Command
Description
Parameter limits
:CALibration
:PROTected
:DC
:ZERO
:V2 <NRf>
:V20 <NRf>
:OHM1M <NRf>
:OHM200K <NRf>
:OHM20K <NRf>
:OHM2K <NRf>
:OHM200 <NRf>
:OHM20 <NRf>
:A200U <NRf>
:A2M <NRf>
:A20M <NRf>
:A200M <NRf>
:A2 <NRf>
:OPEN
User calibration subsystem.
Short-circuit calibration
+2V DC calibration step
+20V DC calibration step.
1M
Ω
calibration step.
200k
Ω
calibration step.
20k
Ω
calibration step.
2k
Ω
calibration step.
200
Ω
calibration step.
20
Ω
calibration step.
200µA DC calibration step.
2mA DC calibration step.
20mA DC calibration step.
200mA DC calibration step.
2A DC calibration step.
Open-circuit calibration step.
0.95 to 2.05
9.5 to 20.5
475E3 to 1.025E6
95E3 to 205E3
9.5E3 to 20.5E3
950 to 2.05E3
95 to 205
9.5 to 20.5
95E-6 to 205E-6
0.95E-3 to 2.05E-3
9.5E-3 to 20.5E-3
95E-3 to 205E-3
0.95 to 2.05
Содержание 2002
Страница 1: ...Model 2002Multimeter Calibration Manual A G R E A T E R M E A S U R E O F C O N F I D E N C E...
Страница 9: ...iv...
Страница 11: ...vi...
Страница 55: ...Calibration 2 26...
Страница 70: ...Specifications A 1 A Specifications...
Страница 83: ...A 14 Specifications...
Страница 101: ...Calibration Programs B 18...
Страница 113: ......