![Keithley 2002 Скачать руководство пользователя страница 25](http://html1.mh-extra.com/html/keithley/2002/2002_calibration-manual_3975661025.webp)
Performance Verification
1-14
8. Set the calibrator output to 190
Ω
, and allow the reading
to settle.
9. Verify that the reading is within the limits stated in Table
1-8. (NOTE: Recalculate limits if calibrator resistance is
not exactly as listed.)
10. Repeat steps 8 and 9 for the 2k
Ω
through 2M
Ω
ranges
using the values listed in Table 1-8. (Do not use offset
compensation for the 200k
Ω
and 2M
Ω
ranges.)
20M
Ω
and 200M
Ω
range verification
1. Connect the DC calibrator and Model 2002 using the 2-
wire connections shown in Figure 1-7.
2. Set the calibrator to the 2-wire mode (external sense
off).
3. Set Model 2002 operating modes as follows:
A. From normal display, press CONFIG then
Ω
2.
B. Select SPEED, then press ENTER.
C. Select HIACCURACY, then press ENTER.
D. Select FILTER, then press ENTER.
E. Select AVERAGING, then press ENTER.
F. Using the cursor and range keys, set the averaging
parameter to 10 readings, then press ENTER.
G. Press EXIT to return to normal display.
4. Select the Model 2002
Ω
2 function, and change to the
20M
Ω
range. (If the FILT annunciator is off, press the
FILTER key to enable the filter.)
5. Set the calibrator to output 19M
Ω
, and allow the reading
to settle.
6. Verify that the reading is within the limits for the 20M
Ω
range stated in Table 1-8. (NOTE: Recalculate limits if
actual calibrator resistance differs from value shown.)
7. Repeat steps 4 through 6 for the 200M
Ω
range (output
100M
Ω
).
Figure 1-6
Connections for resistance verification (20
Ω
-2M
Ω
ranges)
2002 MULTIMETER
Model 2002
5700A Calibrator (Output 4-wire Resistance)
Input HI
Output HI
Input
LO
Output
LO
Note
: Use shielded cables to minimize noise.
Enable calibrator external sense mode.
Sense HI
Sense LO
Sense HI
Sense LO
1.90000000 k
Ω
OCmp
Figure 1-7
Connections for resistance verification (20M
Ω
and 200M
Ω
ranges)
2002 MULTIMETER
Input HI
Output HI
Input
LO
Output
LO
Model 2002
5700A Calibrator (Output 2-Wire Resistance)
Note
: Use shielded cable to minimize noise.
Disable calibrator external sense mode.
19.0000000 M
Ω
Содержание 2002
Страница 1: ...Model 2002Multimeter Calibration Manual A G R E A T E R M E A S U R E O F C O N F I D E N C E...
Страница 9: ...iv...
Страница 11: ...vi...
Страница 55: ...Calibration 2 26...
Страница 70: ...Specifications A 1 A Specifications...
Страница 83: ...A 14 Specifications...
Страница 101: ...Calibration Programs B 18...
Страница 113: ......