4-12
8. Start scanning as follows:
A. Press SCAN, enter a SCAN COUNT of 99999, then
press ENTER.
B. Select SCAN TIMER operation, then press EN-
TER.
C. Choose a scan INTRVL of 1.5, then press ENTER.
Sense
Ω
LO
Thermistor
Probe
Input HI
Input LO
OUT A HI
OUT A LO
Copper Wires
Model 2001 #1
(Measure 4-wire ohms)
NEXT
DISPLAY
PREV
POWER
DCV
ACV
DCI
ACI
Ω
2
Ω
4
FREQ
TEMP
REL
TRIG
STORE
RECALL
INFO
LOCAL
CHAN
SCAN
CONFIG
MENU
EXIT
ENTER
RANGE
AUTO
FILTER
MATH
RANGE
2001 MULTIMETER
SENSE
Ω
4 WIRE
HI
INPUT
LO
INPUTS
CAL
500V
PEAK
F
R
FRONT/REAR
2A 250V
AMPS
350V
PEAK
1100V
PEAK
Sense
Ω
HI
Input HI
Input LO
Model 2001 #2
(Measure DCV)
R151 CAL
adjustment
Note : Install card in Model 2001 #2
Figure 4-7
Calibration with thermistor probe
D. Choose NO DATA TO MEMORY, then press
ENTER.
E. Press ENTER to begin scanning.
9. Note the reading on the Model 2001 display when Chan-
nel 2 is closed. If the display reads other than 32.0
°
F, ad-
just trimmer R151 on the Model 2001-TCSCAN card
until the Model 2001 reads 32.0
°
F,
±
1 count.
10. Press EXIT on the Model 2001 to stop the scan.