AC calibration commands
The :AC commands perform comprehensive (user) calibration of the ACV and ACI functions.
Table 3-3 summarizes these comprehensive calibration commands.
Table 3-3
AC calibration commands
Command
Description
:CALibration
:PROTected
:AC
:STEP
1
2
3
4
5
6
7
8
9
10
11
12
13
10mV AC at 1kHz calibration step.
100mV AC at 1kHz calibration step.
100mV AC at 50kHz calibration step.
1V AC at 1 kHz calibration step.
1V AC at 50kHz calibration step.
10V AC at 1kHz calibration step.
10V AC at 50kHz calibration step.
100V AC at 1kHz calibration step.
100V AC at 50khz calibration step.
100V AC at 1kHz calibration step.
100mA AC at 1kHz calibration step.
1A AC at 1kHz calibration step.
2A AC at 1kHz calibration step.
Calibration Command Reference
3-15