![Keithley 2000 Скачать руководство пользователя страница 39](http://html.mh-extra.com/html/keithley/2000/2000_calibration-manual_661840039.webp)
DC current calibration
After the 1M
Ω
resistance point has been calibrated, follow these steps for DC current cali-
bration:
1.
Connect the calibrator to the AMPS and INPUT LO terminals of the Model 2000 as
shown in Figure 2-3.
2.
Perform the calibration steps listed in Table 2-10. For each step:
• Set the calibrator to the indicated current, and make sure the unit is in operate. (Use
the recommended current if possible.)
• Send the indicated programming command. (Change the current parameter if you are
using a different calibration current.)
• Wait until the Model 2000 completes each step before continuing.
NOTE
If you are performing DC-only calibration, proceed to the “Programming calibration
dates” paragraph.
Table 2-10
DC current calibration programming steps
Calibration
step
Calibrator
current
Calibration command*
Parameter
range
10mA
100mA
1A
10.00000mA
100.00000mA
1.000000A
:CAL:PROT:DC:STEP10 10E-3
:CAL:PROT:DC:STEP11 100E-3
:CAL:PROT:DC:STEP12 1
9E-3 to 11E-3
90E-3 to 110E-3
0.9 to 1.1
* Change parameter if using a different current.
Calibration
2-17
Содержание 2000
Страница 1: ...Model 2000Multimeter Calibration Manual A G R E A T E R M E A S U R E O F C O N F I D E N C E ...
Страница 10: ...1 Performance Verification ...
Страница 23: ...2 Calibration ...
Страница 45: ...3 Calibration CommandReference ...
Страница 65: ...A Specifications ...
Страница 76: ...A 12 Specifications ...
Страница 77: ...B Error Messages ...
Страница 80: ...C Calibration Program ...